Entering Vent Maintenance Mode

WARNING
®
Ventilator Checkout and Maintenance Modes - The LTV
1200, 1150, and 1100 ventilator
does not deliver gas during the ventilator Checkout mode (VENT CHECK) or ventilator
Maintenance mode (VENT MTNCE) and should not be used to ventilate a patient during these
tests.
The Ventilator Maintenance menu is not enabled when the ventilator is powered up
normally. To enable the Ventilator Maintenance menu:
1) Turn the ventilator off.
2) Place the ventilator in an ESD-safe environment, and use ESD controls.
3) Remove the back panel of the ventilator. For instructions see Chapter 8 - Component
Removal and Replacement.
4) Locate DIP switch #5 and set it to ON (see illustration)
5) Turn the ventilator on.
®
p/n 18603-001, Rev. J