CareFusion LTV 1200 Service Manual page 191

Table of Contents

Advertisement

Flow Valve preparation:
1) If replacing the same Flow Valve, skip to the next section.
2) If installing a new Flow Valve, P/N 10019, measure the length of the clear silicone tube
attached to the port above the orange silicone tube (see illustration).
new Flow Valve assemblies are shipped with 6.00" clear silicone tubing to
accommodate use on most LTV
For use with LTV
length of 5.00"  0.250" prior to installing the new Flow Valve
3) New Flow Valve assemblies are also shipped with Spiral Wrapping on the 4-wire
connector cable. For use with LTV
must be gently removed prior to installing the Flow Valve (unwrap, do not pull on, or
strain the wiring).
do not remove the Warning label from the 4-wire connector cable
Page 8-70
®
ventilators
®
1200, 1150, and 1100 ventilators, the tube must be cut to a total
®
1200, 1150, and 1100 ventilators, the Spiral Wrap
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents