Solenoid Mount Assembly - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Solenoid Mount Assembly

Parts required for replacement:
Solenoid Mount Assy. P/N 18607-001
Replace if damaged:
(4-24 X 3/8 Pan Head ) Screw P/N 17682-103
(.125 I.D. X .3125 O.D.) Washer P/N 18032-001
NOTE
Prior to disconnecting or removing any of the tubes, note their positions and review the tube
routing configurations tables, instructions and diagrams beginning on page 8-33.
To remove the solenoid mount assembly:
1) Disconnect the 4-wire flow valve connector from the motor board.
2) Disconnect the Pisco connector from the oxygen pressure transducer of the analog
®
board. (LTV
1200 only) See instructions on page 8-57.
3) Disconnect the 3-wire flow valve connector from the power board.
4) Remove the flow valve assembly (see instructions on page 8-66).
5) Disconnect the 2-wire connectors from the purge solenoids.
Page 8-126
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with
torque meter
Grounded anti-static
wrist strap
120 in-oz Torque
Screwdriver
p/n 18603-001, Rev. J