CareFusion LTV 1200 Service Manual page 172

Table of Contents

Advertisement

4)
Attach the Pisco connector onto the O
PCBA, as shown (LTV
5)
Insert the tube from the O
least 1/2"), as shown (LTV
gently pull the Pisco connector (~2-lbs force) while holding the Analog
PCBA in place, then pull the tube while holding the Pisco connector to
ensure they are both properly attached
6)
Route the 4-wire Flow Valve cable over previously installed tubing and attach the
connector to the 4-wire connector (J4) on the Motor PCBA, as shown. The
connector is keyed to fit only one way and will snap into place when properly aligned.
for clarity, a warning label ("WARNING, DO NOT DISCONNECT WHILE
UNIT IS ON!") attached to the 4-wire cable is intentionally not shown in the
following illustration
7)
Route the orange tube (with elbow) from the Flow Valve over previously installed
tubing and onto the Valve Differential Transducer Low Pressure port, as shown.
p/n 18603-001, Rev. J
®
1200 only).
Blender straight into the top of the Pisco connector (at
2
®
1200 only).
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Pressure Transducer port on the Analog
2
Page 8-51