Turbine Manifold

Parts required for replacement:
Turbine Manifold Assembly P/N 11490
Cable Tie P/N 10466
O
Donut Seal P/N 10603
2
Silicone Gel Lubricant P/N 10123
RTV Silicone Adhesive P/N 10122
Replace if damaged:
Thermo Conductive Pad P/N 10129
Sealing Gasket P/N 10175
Damping Grommets (4) P/N 10266
Grounding Clips (4) P/N 10752
Green O
Port Cover (O
2
When replacing the turbine manifold assembly, it is easiest to remove several components
and replace them together. These will be done in the following order:
Remove the back panel and disconnect the battery
Remove the left soft side
Remove the motor board
Remove the flow valve
Remove the turbine manifold and oxygen blender (LTV
and 1100) as a unit
Disconnect the oxygen blender (LTV
turbine manifold
Connect the new oxygen blender (LTV
turbine manifold
Replace the oxygen blender (LTV
manifold as a unit
Replace the flow valve
Replace the motor board
Replace the left soft side
Reconnect the battery and replace the back panel
70
In the European Union, Loctite® 8104 may be substituted as an equivalent compound.
p/n 18603-001, Rev. J
70
Block) P/N 14446
2
®
1200) or O
®
1200 ) or O
®
1200 ) or O
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with torque
meter
Grounded anti-static wrist strap
Small dykes or cutters
Mild cleanser
Cable tie tool
Flow Valve Insertion Tool (Mylar)
P/N 14206
Torque screwdriver
(120 in-oz/o.84 Nm)
®
1200) or O
2
®
Block (LTV
1150, and 1100) from the
2
®
Block (LTV
1150, and 1100) to the
2
®
Block (LTV
1150, and 1100) and turbine
2
®
Block (LTV
1150,
Page 8-131