Tubing Removal/Replacement Instructions - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Tubing Removal/Replacement Instructions

Parts required for replacement:
None
Replace if damaged:
46
Flexible tubing
Elbow fitting (1) P/N 11592
Elbow fitting (2) P/N 10958
Tee fitting (3) P/N 18186-001
Wye fitting (1) P/N 18185-001
Plug fitting (1) P/N 18187-001
Restrictor (1) P/N 18551-022
Filter (1) P/N 18552-001
Cable assembly (1) P/N 18530-001
To remove internal flexible tubing:
1)
Remove the ventilator back panel and disconnect the internal battery cable from
the Power PCBA (see instructions on page 8-28).
2)
Review and become familiar with the existing tubing configuration prior to
removing individual tubes and strictly adhere to the replacement instructions when
reattaching them (point to point and layering).
3)
Remove only the tubing or components that needs replaced, upgraded, or removed
in order to access other components requiring service.
all tubing is interference fit onto individual ports, elbows or tees and can be
removed by gently twisting the tube and pulling it off.
4)
To avoid damaging tubing and
expedite replacement;
do not remove the Tube
and Tee subassembly
attached to solenoid
manifold port # 9 (side
port) until either the
Analog PCBA or Solenoid
Manifold is removed and
then only when
necessary to replace
individual tubes or
components of the
subassembly itself
46
See Tubing Table in this chapter for detailed descriptions and part numbers of individual flexible tubes.
p/n 18603-001, Rev. J
Tools required:
Grounded anti-static wrist strap
Measuring tool (Calipers, Ruler, etc) 0" thru 11" range
Pliers - long round nose, smooth jaws – ESD safe
Torque Driver(s) – 20 in-oz thru 120 in-oz range with;
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
# 1 Phillips Bit
1/4" Nut Driver
Page 8-37