Alarm Test

The alarm test verifies that the audible alarm is working correctly.
To run the alarm test:
1) Press Select while ALARM is displayed.
2) Verify that the audible alarm sounds. When the alarm has sounded for at least 2
seconds, press Select again.
 The audible alarm is silenced and the next menu item is displayed.
3) Verify a confirming audible chirp sounds after the alarm is silenced.
If the ventilator fails the alarm test, see Chapter 7 - Troubleshooting for more information.
NOTE
®
Although the LTV
1200 Front Panel is shown, the test is applicable to the LTV
1100 Ventilators.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
1150 and
Page 2-3