NOTE
45
An O
Sampling Tube
2
General Checkout testing, must be installed before replacing the ventilator's back panel. See
General Checkout in Chapter 9 – Final Checkout Test for test requirements.
16) Route an O
Sampling Tube as shown in the illustration;
2
through the louver in the side panel,
between the Solenoid Manifold and the Analog PCBA,
under the Flow Differential Transducer,
over the tube attached to Solenoid Manifold port # 9 (side port),
between the Valve Differential Transducer low and high pressure ports,
under the Airway Pressure Transducer port,
under the Pisco connector,
over the Motor PCBA,
beside the Motor PCBA Heatsink and,
on top of the 3-wire Turbine Manifold connector
45
O
Sampling Tube, P/N 10544, ~10.0" long, 0.125" O.D. X 0.079 I.D. clear polycarbonate tubing.
2
p/n 18603-001, Rev. J
used for testing internal oxygen enrichment (oxygen leakage) during
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-31