Caring for the Internal Battery

To preserve maximum battery life:
Fully recharge the Internal Battery every 2 months while the ventilator is in storage.
Recharge the battery by plugging the ventilator into an AC power source for 24
hours. If the LTV
hours, or if it is illuminated red, immediately contact a certified
technician or CareFusion.
Store the ventilator at temperatures less than 60°C (140°F).
A new battery should be charged on standby for 24 hours prior to use on a patient.
CAUTION
Storage Temperature - Storing the LTV
temperatures above 60°C (140°F) for long periods can damage the internal battery and
cause expected battery duration to degrade.
Battery Disposal:
®
The LTV
1200, 1150, and 1100 ventilator's internal battery, P/N 18608-001, is a sealed
lead acid battery. Some jurisdictions consider these batteries hazardous materials subject to
special disposal regulations. Contact the proper agency for information on permissible
methods of disposing of used batteries.
Page 8-84
®
battery Charge Status LED is not illuminated green within 24
®
ventilator and/or the internal battery at
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
service
CareFusion
p/n 18603-001, Rev. J