To connect the Turbine Manifold to a new Oxygen Blender/O
1)
Handle the manifold and blender/O
carefully so as not to break the silicone seal
around the oxygen tube entering the manifold.
If this seal is damaged, repair with RTV
silicone adhesive.
2)
Cut the cable tie around the tube connection
to the oxygen blender/O
oxygen blender/O
manifold.
3)
Make a loose loop with a cable tie and slide it
over the tube that connects the turbine
manifold to the oxygen blender/O
Connect the tube to the barbed fitting on the
oxygen blender/O
4)
Using a cable tie tool, tighten the cable tie to
2 tension. The cable tie tool should trim the
tail off close to the connector.
To install the Turbine Manifold and Oxygen
Blender/O
Block:
2
1)
Slide the turbine manifold and oxygen blender/O
any tubing or wiring under the manifold while it is being installed.
2)
Replace the 4 black screws (P/N
10918B) into the turbine
manifold. Torque to 20 in-oz
(0.14 Nm).
p/n 18603-001, Rev. J
Block
2
Block. Separate the
2
Block and turbine
2
Block.
2
Block.
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Block:
2
Block into place. Be sure not to catch
2
Page 8-99