CareFusion LTV 1200 Service Manual page 298

Table of Contents

Advertisement

Volume
Alarm Volume
Vent Inop
Ventilator Inoperative
Mechanical Controls
Control
Over Pressure Relief
Sub-Ambient Relief
Internal Compliance
Compliance
Monitors
Monitor
Calculated Peak Flow
Exhaled Tidal Volume
I:E Ratio, Measured
I:E Ratio, Calculated
Mean Airway Pressure
O
Cylinder Duration
2
®
(LTV
1200 Only)
Peak Inspiratory
Pressure
PEEP
Total Breath Rate
Total Minute Volume
SBT Minutes
f/Vt
and
f
p/n 18603-001, Rev. J
60 to 80 dBA at one meter
Immediately upon a ventilator inoperative condition, the audible
indicator will begin sounding with a steady tone and the Vent Inop
LED shall illuminate. Depressing the alarm Silence Reset button
will silence the audible indicator.
Range
≤125 cmH
O
2
Pressure Drop:  5 cmH
< 0.1 mL/cm
Range
10 to 100 Lpm
0 to 4000 ml
99:1 and 1:99
Based on the measured
inspiratory / exhalation times
1:99 to 4.0:1 based on set
breath rate and inspiratory
time
0 to 99 cmH
O
2
0 - 99 hours and 59 minutes
0 to 120 cmH
O
2
0 to 99 cmH
O
2
0 to 250 breaths per minute
0 to 99.9 liters
1 – 120 minutes
0 – 4000
0 – 250bpm
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
 5 dBA
Tolerance
N/A
at 50 Lpm
O
2
Tolerance
2 Lpm or  10%, whichever is greater
 15% or 15 ml, whichever is greater
Accuracy for times are  50 ms or 5%,
whichever is greater
 5%
 2 cmH
O or 10%, whichever is greater
2
- 0 / + 40%
 2 cmH
O or 5%, whichever is greater
2
 2 cmH
O or 10%, whichever is greater
2
 1 bpm or within 5% of the breath
period, whichever is greater
 15%, or the measured total breath rate
times 15 ml, whichever is greater
 0.1 seconds
 20%
Total breath rate:  1 bpm or within 5%
of breath period, whichever is greater
Page A-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents