CareFusion LTV 1200 Service Manual page 236

Table of Contents

Advertisement

19) Replace the alarm sounder, (see instructions on page 8-55).
20) Reconnect the 2-wire fan connector, solenoid mount assembly 4-wire connector, 4-wire
rotary switch connector and 2-wire sounder connectors to the power board. The
connectors are keyed to fit in only one direction and will snap into place when properly
connected.
21) Replace the analog board and install two 1/4" Green colored pan-head mounting
screws (P/N 10435G). Torque-tighten to 60 in-oz (0.42 Nm). See page 8-57 for
instructions.
22) Ensure that the flexible tube from port B of the accumulator is tucked between the purge
solenoids and the turbine manifold. Replace the turbine manifold spacer using four (4)
7/16" pan head screws (P/N 10433). Torque tighten to 20 in-oz (0.14 Nm).
23) Connect the purge solenoid cables from the power board to the purge solenoids.
Orientation does not matter.
24) Reattach the solenoid mount assembly to the turbine manifold spacer using the single
(1) screw and washer. See instructions on page 8-126. Torque the screw to 120 in-oz
(0.84Nm).
25) Reattach the accumulator to the turbine manifold spacer using the two (2) 7/16" pan
head screws. See instructions on page 8-53. Torque both screws to 60 in-oz (0.42 Nm).
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-115