4)
Insert and thread two 1/4" flat-head mounting screws with finish washers into the
screw holes in the ventilator's side panels, as shown in the illustration.
Finish washers should be already in place.
5)
Torque tighten the mounting screws to these specified values
Torque tighten the screws in the back panel of the ventilator to 60 in-oz (0.42 Nm)
Torque tighten the two screws in the sides of the ventilator to 20 in-oz (0.14 Nm)
CAUTION
Damage to finish washers -Do not over tighten screws to avoid damage to the finish
washers.
Page 8-8
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J