Analog Board Assembly - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Analog Board Assembly

Parts required for replacement:
Analog PCBA Assembly P/N 18242-001
Replace if damaged:
Pisco connector P/N 10543
To replace the analog board:
1) Remove the back panel of the ventilator and disconnect the internal battery cable (see
instructions on page 8-28).
2) Detach the 6 flexible tubes from the analog board (one orange tube from the flow valve,
one tube from the accumulator and 4 from the solenoid manifold.
NOTE
Tubing Configurations - When disconnecting tubes, note their positions and review the
internal flexible tube routing tables, instructions and diagrams beginning on page 8-33.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with torque meter
Grounded anti-static wrist strap
Page 8-59