5)
Attach subassemblies to the Solenoid Mount assembly, as follows:
attach the open end of the Plug subassembly tube to the bottom port of the
Pilot-in solenoid on the Solenoid Mount P/N 18607-001
attach the end of the Wye subassembly tube with the installed restrictor
onto the middle port of the Pilot-in solenoid (restrictor will slide farther into
tube during attachment)
attach the other top tube of the Wye subassembly onto the bottom port of
the Pilot-out solenoid
attach the end of the Filter subassembly tube with the filter onto the top port
of the Pilot-in solenoid (filter will slide farther into tube during attachment)
attach the open end of the Tee subassembly to the middle port of the Pilot-
out solenoid
attach the Pilot Solenoid/Power Board Cable assembly P/N 18530-001, two
red wire connector onto the Pilot-in solenoid leads and the two black wire
connector onto the Pilot-out solenoid leads
Page 8-42
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J