4)
Construct a Tube and Tee subassembly, as follows;
Insert Tee fitting P/N 18186-001 into 2.06" and 1.75" of tubing P/N 11834
Attach the open end of the 1.75" tubing to manifold side port # 9
5)
Attach the Tee & Elbow subassembly to solenoid manifold port # 1, oriented as
shown, with;
the open end of the tee facing towards the tube on port # 2
the open end of the elbow facing towards the Purge Solenoids
6)
Attach the Tube and Tee subassembly to solenoid manifold port # 9, so that the
open end of the tee is oriented as shown when the tubing is bowed over as shown.
7)
As applicable, install the Solenoid Manifold onto the Power PCBA (see instructions
on page 8-120).
Page 8-40
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J

Hide quick links: