CareFusion LTV 1200 Service Manual page 168

Table of Contents

Advertisement

6)
Connect the open end of the Wye subassembly tubing onto the tee fitting in the
Tube and Tee subassembly attached to Solenoid Manifold port # 9.
7)
Connect the open end of the Filter subassembly onto the tee fitting in the Tee and
Elbow subassembly attached to Solenoid Manifold port # 1.
8)
Route the Tee subassembly between the tubes on Solenoid Manifold ports # 3 and
# 4, under the elbow fitting on port # 7 and between Solenoid Manifold ports # 6
and # 7.
the tee on the end of the tube must be positioned at the opening of the
Louver to the outside of the LTV
9)
Insert the elbow fitting attached to the tubing on Solenoid Manifold port # 7 into the
tubing attached to the Flow Differential Transducer on the Analog PCBA.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
Page 8-47