6) Connect the male Luer fitting from the calibration syringe assembly to the high pressure
female Luer fitting on the side of the ventilator. Decrease and maintain the vacuum until
a value in the range of 10 to 150 is displayed in the Tidal Volume window. If necessary,
add a large amount of compliance, such as a test lung, to the calibration syringe
assembly, this will significantly stabilize the readings during this step.
During this step, any value in the range of 10 to 150 displayed in the Tidal Volume
window is equally acceptable, but it is very important that the value displayed in the
Tidal Volume window is stable when pressing Select.
Alternatively, it is also acceptable to apply a small positive pressure to the ventilator low
pressure fitting to fulfill this step.
Record the number that was displayed when you pressed Select on the
Calibration Worksheet.FD 0
7) Disconnect from the ventilator so the connection is open to ambient room air. Observe
the numbers displayed in the Tidal Volume setting window. When the displayed value is
stable, press Select.
Record the number that was displayed when you pressed Select on the Calibration
Worksheet.
FD 4015 ± 70 AD is displayed.
8) Connect the male Luer fitting from the Calibration Syringe assembly to the high pressure
female Luer fitting. Increase and maintain the pressure until a value of 3945 to 4085 is
displayed in the Tidal Volume window. If necessary, add a large amount of compliance,
such as a test lung to the Calibration Syringe assembly, this will significantly stabilize the
readings during this step.
During this step, any value in the range of 3945 to 4085 displayed in the Tidal Volume
window is equally acceptable, but it is very important that the value displayed in the
Tidal Volume window is stable when pressing Select.
Record the number that was displayed when you pressed Select on the Calibration
Worksheet.
C
FD 0
M
9) Disconnect from the ventilator so the connection is open to ambient room air. Observe
the numbers displayed in the Tidal Volume setting window. When the displayed value is
stable, press Select.
Record the number that was displayed when you pressed Select on the Calibration
Worksheet.
VALVE DIFF is displayed if the calibration was successful.
If CAL FAIL appears at any time, press Select to exit the calibration. When FLOW DIFF is
displayed, press Select and perform the calibration again.
p/n 18603-001, Rev. J
C
H
O is displayed.
M
2
H
O is displayed
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 6-13