Cleaning or Replacing the O
The O
inlet filter should be cleaned or replaced when it becomes soiled. Failure to do this
2
can affect ventilator performance.
CAUTION
Oxygen Supply Contamination - The accuracy of the oxygen delivery capabilities of LTV
Ventilator can be compromised by foreign debris contamination in the oxygen supply
system. To reduce the risk of airborne contaminants entering the ventilator, ensure that any
oxygen supply connected to the ventilator is clean, properly filtered
O
inlet port cap is securely installed on the O
2
connected to an external oxygen supply.
To clean or replace the O
1) Disconnect the high pressure O
ventilator.
2) Using a pick, gently remove the rubber O-Ring from inside the O
Do not damage the O-Ring while removing it. Tip the ventilator to allow the O
to slide out.
3) Clean the filter using a mild cleanser, warm water and a soft brush. Rinse the filter
thoroughly to remove all traces of the cleanser. Allow the filter to dry completely before
replacing it in the ventilator.
4) Inspect the filter for damage. If the filter is not intact, shows signs of damage or cannot
be completely cleaned, replace it with a new O
001) and O-Ring (P/N 10609), available from CareFusion.
7
In addition to the existing internal O
(P/N 14470) is available from CareFusion.
Page 4-4
Inlet Filter (LTV
2
Inlet Filter:
2
hose from the oxygen block on the left side of the
2
Inlet filter, P/N 14313 or P/N 19845-001, an External, In-Line Oxygen Filter
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
7
Inlet Port whenever the ventilator is not
2
Inlet Filter (P/N 14313 or P/N 19845-
2
1200 Only)
and that the ventilator's
inlet port. Use caution:
2
Inlet Filter
2
p/n 18603-001, Rev. J
®