Solenoid Manifold Preparation:
1)
Attach tubing to the Solenoid Manifold P/N 18528-001 ports;
Port # 2 – 4.70" of tubing P/N 10458
Port # 3 – 3.25" of tubing P/N 11834
Port # 4 – 1.75" of tubing P/N 11834
Port # 6 – 2.40" of tubing P/N 11834
Port # 8 – 2.40" of tubing P/N 11834
2)
Insert the smaller end (O.D.) of Elbow fitting P/N 11592 into 0.62" of tubing P/N
11834 and attach the tubing to manifold port # 7.
3)
Construct a Tee & Elbow subassembly, as follows;
Insert Tee fitting P/N 18186-
001 into 1.10" and 0.62" of
tubing P/N 11834
Insert the larger end (O.D.) of
the Elbow fitting P/N 10958
into the open end of the 0.62"
of tubing (ensure that the
open ends of the elbow and
tee fittings are rotated 90°
from each other, as shown)
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-39