Remove or loosen the one (1) screw with washer that secures the solenoid mount
assembly to the spacer.
9)
Remove the two (2) screws attaching the accumulator to the spacer and remove the
accumulator from the turbine manifold spacer.
10) Remove the four (4) screws that secure the turbine manifold spacer to the turbine
manifold. Carefully remove the turbine manifold spacer and set it aside, away from the
ventilator.
11) Remove the analog board (see instructions on page 8-57).
12) Disconnect the 4-wire solenoid mount connector from JP16 on the power board.
Page 8-106
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J