Power (cont.)
Internal Battery
12V sealed lead acid battery. 4.5Ah.
Feature
Full Power
Medium Power
Low Power
Charge Time
Charge Status
Hysteresis
Minimum battery duration following
eight (8) hour charge to ventilator
shutdown
Minimum time from battery full (green
LED) to battery low (amber LED and
BAT LOW alarm):
Minimum time from battery low to
battery empty (red LED and BAT
EMPTY alarm):
Minimum time from battery empty to
"ventilator shutdown" (Vent Inop LED
and Vent Inop alarm):
DOT Requirements:
Agency Requirements
Regulatory Requirements
FDA Draft Reviewer Guidance for Ventilators, July, 1995.
Shipping Requirements
The ventilator, packed in its shipping container, conforms to the International Safe Transit
Association requirements for packaged products weighing less than 100 pounds.
Page A-10
Range
Green LED
Amber LED
Red LED
Battery shall be capable of being >90%
charged within 8 hours, from fully
discharged state to state indicated by
green Charge Status LED.
Pre-Charge Qualification:
Battery Charging:
Battery >90% Charged:
Battery Fault:
Ventilator shall not resume battery
operation unless the battery voltage
level is 11.8 V.
Unregulated, meets the requirements of 49 CFR 173, 159 (d).
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tolerance / Indicators
When external power is
present, and the vent is
running at the nominal load
Flashing Amber LED
Amber LED
Green LED
Red LED
 2%
40 minutes
Nominal Load Settings:
28 minutes
Mode
PEEP
Breath Rate (bpm)
O
%
2
8 minutes
Tidal Volume (ml)
Lung Compliance (ml/cmH
Insp. Time (sec)
ET Resistance (cmH
4 minutes
Sensitivity (Lpm)
Battery Temp.
A/C
5
15
21
800
O) 50
2
1.5
O/L/S)
5.87
2
2
o
25
C
p/n 18603-001, Rev. J