TEST
DESCRIPTION
Alarm Settings
Low O
Pres alarm (LTV
2
Set O
% to 22% (LTV
2
Low Minute Volume alarm
LMV alarm set to 10
Low Pressure alarm
LP alarm set to 60 cmH
High Pressure alarm
HP alarm 10 cmH
O below
2
monitored PIP
®
O
Enrichment (LTV
1200 only)
2
O
Inlet pressure @ 50 ± 2 PSIG,
2
%O
@ 60%
2
Disc / Sense Alarm
High side sense line disconnected
from unit
Pressure Operation
(LTV 1200, 1150 only)
Control Settings:
Pressure Assist / Control
Breath Rate = 10
Pressure Control = 45
Insp. Time = 1.5
PEEP = 5
Rise Time = As required
Power Alarm
Return to normal ventilation mode
(LTV 1200, 1150 only).
Disconnect AC adapter from ventilator
Vent Inop Alarm
Clear the POWER LOST alarm. Turn
the ventilator off. Wait 15 seconds.
Clear the Vent Inop alarm. Test is
complete.
80
Set the Rise Time under the VENT OP menu as required (1 through 9, depending on test lung used), to
achieve 50 cmH
O of pressure.
2
Page 9-10
STEP
#
®
1200 only)
4)B)
®
1200 only)
4)C)
4)D)
O
2
4)E)
4)G)
4)I)
4)K)
80
4)L)
4)M)
4)N)
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
MEAS.
REQUIREMENT
VALUE
Alarm must activate immediately
Alarm must activate at start of next
breath
Alarm must activate at end of the next
inspiration
Alarm must activate on the next breath
External O
monitor = 55% to 65%.
2
No alarms activate.
Alarm must activate at start of next
inspiration
Monitors should read as follows:
PIP:
No alarms activate.
POWER LOST alarm activates.
Battery Level LED illuminates showing
the charge level.
Ventilator continues to operate from
internal battery
After the ventilator is turned off, the Vent
Inop alarm sounds continuously for a
minimum duration of 15 seconds.
Vent Inop LED illuminates continuously
for a minimum duration of 15 seconds.
Confirming audible chirp must activate
after alarm is silenced
46 to 54 cmH
O
2
p/n 18603-001, Rev. J
PASS
(P) /
FAIL
(F)

Hide quick links: