CareFusion LTV 1200 Service Manual page 300

Table of Contents

Advertisement

Displays
Display
Airway Pressure
Display Window
Patient Effort
Vent Inop
External Power
Charge Status
Battery Level
Usage Meter
Usage Meter
Packaging
Size
Weight
Sound Level
Sound Level
Storage and Operating Conditions
Specification
97
Storage
Temperature
Humidity
Operating
Temperature
Humidity
Orientation
The ventilator functions within its performance specifications when operated in any orientation.
Inlet Air Filtration
The ventilator air filter is removable and cleanable by the operator. All filter materials are FDA
compliant for breathing circuits and meet burn requirements for UL 94HB.
97
®
LTV
ventilators stored at temperatures outside of the specified Operating Temperature range are to be
allowed to stabilize to within the operating temperature range before turning the ventilator on.
p/n 18603-001, Rev. J
Range
-10 to 108 cmH
O
2
12 characters
Green LED
Red LED
Amber / Green LED
Red / Amber / Green LED
Red / Amber / Green LED
1 to 139,000 hrs
3" x 10" x 12" -OR- 3.25" x 10.5" x 13.5" with Protective Boots installed.
13.4 lbs -OR- 14.4 lbs with Protective Boots installed.
Shall not exceed 50 dBA (RMS) at one meter
-20 to 60 degrees C
10% to 95% Relative, non-condensing
+5 to 40 degrees C
15% to 95% Relative, non-condensing
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tolerance
 3 cmH
O or 5%, whichever is
2
greater
n/a
n/a
n/a
n/a
n/a
n/a
Below 100 hrs:  10%
100 hrs:  5%
Tolerance
n/a
n/a
n/a
n/a
Above
Page A-7