Control Test
At the end of the display test, CONTROL shows in the LED display window. The control test
is used to verify that the ventilator buttons and the Set Value knob are working correctly.
To run the control test:
1)
Press Select while CONTROL is displayed.
2)
SELECT is displayed in the display window.
NOTE
®
Although the LTV
1200 Front Panel is shown, the test is applicable to the LTV
1100 Ventilators.
Page 2-6
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
1150 and
p/n 18603-001, Rev. J