Power
Feature
Input Voltage
External Power
AC Adapter
Full Power
Low Power
External Power Off
Hysteresis
Nominal Current
Draw
Nominal Power
Draw
Leakage Current
Ground Resistance
Dielectric Strength
p/n 18603-001, Rev. J
Range
11 to 15 VDC
Input: 100 to 250 VAC, 50 to 60
Hz
Output: 13 VDC
Voltage  11.5 V
Voltage < 11.0V and  9.5V
Voltage < 9.5V
Ventilator shall not resume
external power operation unless
voltage is 11.5V
Startup:
Running:
Startup:
Running:
Total leakage current to Earth ground for the ventilator with only
approved accessories attached, shall not exceed 500 microAmps
during normal operation, per IEC 601-1.
Total leakage current to Earth ground for the ventilator shall not exceed
one milliAmp when any single fault condition is present, per IEC 601-1.
Total impedance between the ground contact at the inlet power
connector and any accessible metal part shall not exceed 0.1 ohm, per
IEC 601-1.
The ventilator shall be able to survive 1500 volts applied from either
phase of the AC power inlet to Earth ground for a period of one minute,
per IEC 601-1.
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tolerance / Indicators
 2.5%
 2%
 2%
 2%
 2%
5.5 amps
3-4 amps
66 watts
36 - 48 watts
Green LED
Amber LED
LED off, switch to
battery
Page A-9