9)
Remove the two (2) screws that secure the accumulator to the turbine manifold spacer,
and remove the accumulator from the spacer.
10) Remove the four (4) screws that secure the turbine manifold spacer to the turbine
manifold assembly.
11) Remove or loosen the one (1) screw that secures the solenoid mount assembly to the
turbine manifold spacer, remove the solenoid mount assembly from the spacer, and set
spacer aside, away from the ventilator.
NOTE
The analog board, solenoid manifold and the solenoid mount assembly cable may be left
connected to the power board.
12) Remove the power board (see instructions on page 8-105). Take care not to catch the
power board on the sounder bracket while removing it.
13) Slide the ZIF connector on the keypad ribbon cable open and carefully remove the
ribbon cable from the connector.
CAUTION
Contact damage - Handle the ribbon cable carefully to prevent any damage to the silver
contact area. Attempting to remove the ribbon cable without opening the ZIF connector will
damage the ribbon cable and may require replacing the front membrane panel.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-87