CareFusion LTV 1200 Service Manual page 248

Table of Contents

Advertisement

6) Loosen or remove the one (1) screw that secures the solenoid mount assembly to the
turbine manifold spacer, and remove the solenoid mount assembly from the spacer.
NOTE
When removing the tubes from the pilot-in solenoid, make sure that the small brass filter and
restrictor remain in the tubes.
7) Remove the three (3) tubes from the pilot-in solenoid, and the two (2) tubes from the
pilot-out solenoid.
8) Disconnect the two (2) wire connectors from the pilot-in and pilot-out solenoids.
To replace the solenoid mount assembly:
1) Reconnect the 2-wire connector (red wires) to the pilot-in solenoid, and the 2-wire
connector (black wires) to the pilot-out solenoid.
2) Reconnect the flexible tubes to the pilot-in and pilot-out solenoids following the internal
flexible tube routing tables, instructions and diagrams beginning on page 8-33.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-127