Event Name
Code
STACK
56
SYNCER0
37
SYNCER1
36
TBN ESTP
46
TBN HSTP
66
TBN ISTP
44
TBN ZERO
45
VENT 0
02
VENT 1
01
VENT CHK
04
WDOG TST
59
XDC FLT0
20
XDC FLT1
19
XDCR AIR
64
XDCR BI
63
XDCR NAR
61
XDCR WID
62
55
N/A
30
N/A
N/A
29
N/A
07
Page D-8
Event
Stack overflow detected
Stepper motor lost sync exited
Stepper motor lost sync occurred
Turbine emergency stop occurred
Turbine Hold Stop occurred
Turbine immediate stop occurred
Turbine set to zero flow occurred
Power off
Power on
Set vent check
Watchdog test run
XDCR fault exited
XDCR fault occurred
Airway pressure transducer fault
Differential pressure transducer -
Bi-directional channel fault
Differential pressure transducer -
Narrow channel fault
Differential pressure transducer -
Wide channel fault
Not used
Not used
Not used
Not used
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Associated Alarm
RESET
HW FAULT
HW FAULT
HIGH PRES
HIGH PRES
HIGH PRES
HIGH PRES
None
None
Entered VENT CHECK mode
Vent Inop
XDCR FAULT
XDCR FAULT
XDC FLT1
XDC FLT1
XDC FLT1
XDC FLT1