Cleaning The Ventilator - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Chapter 4 - C
, D
S
LEANING
ISINFECTING AND
TERILIZING

Cleaning the Ventilator

All ventilator external surfaces should be cleaned before and after each patient use, and as
otherwise necessary.
To clean the ventilator:
1) Wipe the exterior surfaces of the ventilator with a clean, damp cloth. The use of an anti-
bacterial cleaning solution is recommended. Be sure to wipe away any residual cleaner.
CAUTION
®
Ventilator Sterilization – To avoid irreparable damage to the LTV
Ventilator, do not
attempt to sterilize it.
Cleaning Agents – To avoid damaging the ventilator's plastic components and front panel,
do not use cleaning agents containing ammonium chloride, other chloride compounds, more
than 2% glutaraldehyde, phenols, or abrasive cleaners.
Ventilator Immersion - Do not immerse the ventilator in liquids.
Exhalation Valve Cleaning - Do not pour or spray liquid cleaners into the exhalation valve.
Front Panel Cleaning – Do not pour or spray liquid cleaners onto the front panel.
®
p/n 18603-001, Rev. J
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 4-1