Solenoid Mount Preparation:
1)
Construct a Plug subassembly, as follows:
fully insert Plug fitting P/N 18187-001 into 0.85" tubing P/N 10456
2)
Construct a Wye subassembly, as follows:
attach a 1.00" tubing P/N 10456 to each top end (2) of Wye fitting P/N
18185-001 and a 0.50" tubing P/N 10456 to the bottom end
insert Restrictor P/N 18551-022 (red in center) up to a maximum of 0.50"
into either one of the top tubes (big end inserted last)
3)
Construct a Filter subassembly, as follows:
insert Filter P/N 18552-001 (screen in center) up to a maximum of 0.50" into
a 1.60" tubing P/N 10456 (tapered end inserted first)
4)
Construct a Tee subassembly, as follows:
fully insert Tee fitting P/N 18186-001 into 1.75" tubing P/N 10456
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-41