Alarm Sounder Assembly (Previous Version) - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Alarm Sounder Assembly (previous version)

Parts required for replacement:
Sounder Bracket P/N 10119
Sounder Assembly P/N 17432-001
Replace if damaged:
Alarm Bumper P/N 10573
1/4" Pan-head Screw (1) P/N 10435
To replace the Sounder Assembly:
1) Remove the ventilator back panel and disconnect the internal battery cable from the
power board (see instructions on page 8-28). Remove the left and right soft side panels.
2) Disconnect the 2-wire sounder connector from the power board.
3) Remove the 1/4" pan-head screw from the sounder bracket, as shown.
4) Remove the sounder bracket.
5) Remove the sounder assembly and alarm bumper.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with
torque meter
Grounded anti-static wrist
strap
Page 8-55