3)
Align and lower the Solenoid Mount assembly into the cavity between the Solenoid
Manifold and the Flow Valve assembly.
as the assembly is lowered into position, loop and position the excess Pilot
Solenoid / Power PCBA cable as shown in the illustration
4)
Attach the Solenoid Mount assembly to the end of the Manifold to Back Panel
Spacer using the screw (P/N 17682-103, 4-24 X 3/8" pan-head) and washer (P/N
18032-001) on the Solenoid Mount.
5)
Align and hold the Solenoid Mount parallel to the Solenoid Manifold (so that it does
not touch it) and torque tighten the screw to 120 in-oz (0.84Nm).
Page 8-46
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
p/n 18603-001, Rev. J