CareFusion LTV 1200 Service Manual page 271

Table of Contents

Advertisement

®
LTV
SERIAL NUMBER: _____________
TEST
DESCRIPTION
Usage Hours, Date and Time
Verification
Power up ventilator and access
extended features.
Display usage hours.
Set current date.
Set current time.
Turn ventilator off and wait a minimum
of 1 minute
Power up ventilator and access
extended features.
Navigate to VENT OP and press Select.
Display Date
Display Time
Oxygen Enrichment (oxygen leakage)
®
LTV
1200: Connect a 50 ± 2 PSIG
oxygen source to the ventilator.
the O
% setting to 100%.
2
After measurement, return O
21%.
®
LTV
1150 and 1100: Connect a 1.0 to
1.5 PSIG oxygen source to the
ventilator.
Operate for a minimum of 2 minutes.
Extract the tube approximately 0.5 inch
per breath.
Page 9-12
STEP
#
6)A)
6)B)
6)C)
6)D)
6)E)
6)F)
6)G)
6)H)
6)I)
7)A)
Set
setting to
2
7)B)
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
MEAS.
REQUIREMENT
VALUE
The displayed usage hours shall be
greater than or equal to the received
usage hours.
If the displayed date is not current,
then set the current date.
If the displayed time does not match
the current time by ± 10 minutes,
then set the correct time.
The displayed date is the current
date.
The displayed time is the current
time.
Peak % O
measurement taken from
2
inside the unit must be  25%.
PASS
(P) /
FAIL
(F)
p/n 18603-001, Rev. J