General Tools:
The following general tools are required to perform various procedures on the LTV
1100 ventilator;
Air supply, (0-50 cmH
Allen wrench, 7/64"
Amp meter, 10 amp to 60 Hz
 Compressed O
source (0-80 PSI regulator)
2
®
(LTV
1200)
Compressed gas source (0-50 cmH
regulator)
Dental pick
Drill Bit, 3/16"
Drill
Multi-Meter, Digital
Nut drivers, 3/16", 1/4", 1/2" & 13mm adapters
for torque wrench
O
Analyzer (calibrated)
2
O
Supply (0-80 PSIG) (LTV
2
O
Inlet connector (LTV
2
Unique Tools:
The following unique tools are required to perform various procedures on the LTV
the 1100 ventilator and are available from CareFusion separately, or as part of the Maintenance and
Calibration Kit, P/N 11566;
Calibration Syringe Assembly
DISS to NIST fitting for high pressure O
Driver, Torque Dial Indicator
External Battery Test Cable Assembly
Flow Valve Insertion Tool
Internal Battery Test Cable Assembly
Lubricant, Silicone Compound
O
Sampling Tube
2
Patient Assist Cable, Normally Closed
Patient Assist Cable, Normally Open
Pinch Clamp
Power PCB Removal Tool
Service Cable Assembly
Stepper Motor Calibrator
Tube Adapter, 22mm
Turbine Pressure Test Adapter Assembly
107
Pop Rivet tool capable of setting .114" diameter shaft, Pop Rivets.
108
DISS/ NIST adapter required to comply with some international gas fittings (not included in kit, P/N 11566).
109
In the European Union, Loctite® 8104 may be substituted as an equivalent compound.
110
O
Sampling Tube, P/N 10544, ~10.0" long, 0.125" O.D. X 0.079 I.D. clear polycarbonate tubing (not included
2
in kit, P/N 11566).
p/n 18603-001, Rev. J
O)
2
O
2
®
1200)
®
1200)
port
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Pliers, needle nose
107
Pop Rivet tool
Pressure gauge (FSD >80 PSI) (LTV
Pressure manometer (0-100 cmH
Screwdriver, Phillips with torque meter
Screwdriver, Straight tip
Spirometer (calibrated)
Test Lung (2 Liter or other large-compliance
reservoir) or
Test Lungs (2 x 1 liter or greater)
Cable tie tool
Torque Screwdriver, 120 in-oz.
Variable DC voltage source (10 Amp)
Wrist strap, grounded anti-static
Y-connector for test lungs
P/N 11471
108
P/N 10702
P/N 11574
P/N 11474
P/N 14206
P/N 11472
109
P/N 10123
110
P/N 10544
P/N 10779
P/N 10780
P/N 11529
P/N 11599
P/N 11485
P/N 10871
P/N 10696
P/N 11567
®
1200, 1150, and
®
1200)
O)
2
®
1200, 1150, and
Page E-15