LTM/LTV mounting bracket temporary removal (earlier model LTM)

Parts required for replacement:
None
To temporarily remove the LTM/LTV mounting bracket:
1)
Lay the disconnected ventilator on a clean dry surface. Use a 7/64" Allen wrench to
remove the three mounting block screws and mounting block, as shown below.
Retain the screws for reuse when the mounting block is reinstalled.
2)
Use a small Philips screwdriver and remove the five ventilator back panel mounting
screws and LTM/ LTV
the Mounting bracket is reinstalled.
p/n 18603-001, Rev. J
Tools required:
®
mounting bracket, as shown. Retain the screws for reuse when
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Phillips screwdriver with torque meter
7/64" Allen wrench
Page 8-23