Ltm/Ltv ® Mounting Bracket New Installation (Earlier Model Ltm) - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

LTM/LTV
mounting bracket new installation (earlier model LTM)
®
Parts Required for Installation:
®
LTM/ LTV
®
LTM/ LTV
Replacement screws kit, P/N 11149
To install the LTM/LTV
®
If LTV
protective boots are installed on the ventilator, the upper boot must be removed in
order to mount the LTM/ LTV
1)
Lay the disconnected ventilator on a clean dry surface, use a Philips screwdriver and
remove the five back panel mounting screws, as shown in the illustration below.
2)
Orient the ventilator and mounting bracket as shown in the illustration on the next page,
position the bracket on the ventilator and align its screw slots with the corresponding
screw holes in the ventilator back panel.
41
Contained in CareFusion LTM to LTV
42
See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for
additional information.
Page 8-26
mounting block (1), P/N 11146
mounting bracket (1), P/N 11125
®
mounting bracket
®
mounting bracket. See page 8-7 for instructions.
®
Mount Assembly Kit, P/N 11003.
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
41
Phillips screwdriver
41
with torque meter
7/64" Allen wrench
42
:
p/n 18603-001, Rev. J

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents