Before removing the back panel

CAUTION
Opening the ventilator - Always turn the ventilator OFF and remove the external power
before opening the ventilator case or attempting to service the ventilator.
Anti-static precautions - Always wear a grounded anti-static wrist strap when handling the
ventilator with the case open. Electrostatic discharge (ESD) can damage ventilator
electronics.

Before replacing the back panel

Several checks must be made before reinstalling the back panel. Follow the guidelines listed
under Back Panel, Reinstallation in this chapter.

After performing any maintenance

®
After performing maintenance on the LTV
1200, 1150, or 1100, the tests detailed in
Chapter 9 - Final Checkout Test must be run. Some maintenance procedures also require
®
that the LTV
1200, 1150, or 1100 ventilator be partially or completely recalibrated.
CAUTION
Verification of operation - After opening the ventilator and performing any maintenance,
verify proper operation of the ventilator by performing the checks and calibrations
recommended in Chapter 9 - Final Checkout Test.
®