Temperature Compensation

The Temperature Compensation menu is used to set the temperature compensation for the
flow valve ON or OFF.
NOTE
This configuration menu is used for many models of LTV
1100, which always has a Thermistor installed, the setting for Temperature Compensation is
always ON.
CAUTION
Selecting TCOMP OFF on the LTV
low flows and volumes to be delivered. This phenomenon may occur at either end of the
operating temperature range, when the unit is warm or when the unit is cold.
The Configuration menu is set up as follows:
TEMP COMP
TCOMP ON
TCOMP OFF
TCOMP EXIT
If the Temperature Compensation is set to Off, set it to On as follows:
1)
Press Select while TCOMP is displayed in the window.
2)
Rotate the Set Value knob until TCOMP ON is displayed.
3)
Press Select.
4)
Pressing Control Lock exits to the previous menu without making any modifications to
the temperature compensation.
5)
The temperature compensation selection can be viewed in the VENT MTNCE menu
when powered up in Vent Maintenance mode.

Configuration Menu Exit

To exit the configuration menu:
1) Press Select while CONFIG EXIT is displayed.
2) CONFIG is displayed. To access other Vent Maintenance menu options, turn the Set
Value knob to the desired item and press Select.
p/n 18603-001, Rev. J
®
1200, 1150, and 1100 will cause slightly high or slightly
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
®
. For the LTV
1200, 1150, and
Page 6-35