CareFusion LTV 1200 Service Manual page 136

Table of Contents

Advertisement

To install the lower protective boot
1)
Lay the ventilator down (front up) and use a Philips-head screwdriver to remove the four
flat-head mounting screws in the ventilator's side panels, as indicated in the illustration
below.
Do not remove the mating finish washers.
2)
Orient the lower boot to the ventilator as shown in the illustration below. Move the boot
into position on the bottom of the ventilator and align its four screw holes with the
corresponding holes in the ventilator side panels.
Ensure the orientation arrows on the bottom of the boot are aligned "up", as shown
below.
3)
Insert and lightly thread four 1/2" flat-head mounting screws with finish washers
through the screw holes in the sides of the lower boot; as indicated in the illustration
below.
4)
Torque tighten all four screws in the boot to 20 in-oz (0.14 Nm).
CAUTION
Damage to finish washers - Do not over tighten screws to avoid damage to the finish
washers.
40
See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for
additional information.
p/n 18603-001, Rev. J
40
:
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Page 8-15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents