4)
Position and hold the Accumulator above the mounting tabs on the Manifold to
Back Panel Spacer, as shown.
5)
Attach the open end of the Tube and Tee subassembly from Solenoid Manifold port
# 9 onto port A of the Accumulator.
6)
Attach the open end of the 3.70" tube from the PEEP Pilot Pressure Transducer
onto port B of the Accumulator.
7)
Align and lower the Accumulator into position on the spacer mounting tabs.
verify the tubing from ports A and B are not kinked and do not overlap
8)
Insert two 7/16" pan-head screws (P/N 10433) through the Accumulator mounting
tabs and torque tighten to 60 in-oz (0.42 Nm).