To remove the Turbine Manifold and Oxygen blender or O
1)
Remove the back panel and disconnect the battery (see instructions on page 8-28).
2)
Remove the left soft side (see instructions on page 8-117).
3)
Remove the motor board (see instructions on page 8-94).
4)
Disconnect the O
5)
Disconnect the O
the analog board as shown (LTV
6)
Remove the flow valve and disconnect the by-pass tubing from the base of the turbine
manifold (see instructions on page 8-66).
7)
Remove the two (2) screws securing the accumulator to the turbine manifold spacer
and loosen the one (1) screw and washer securing the solenoid mount assembly.
Detach the accumulator and the solenoid mount assembly from the spacer.
Page 8-132
blender 8-wire connector from the power board (LTV
2
tube from the Pisco connector on the oxygen pressure transducer on
2
®
1200 only ) .
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Block:
2
®
1200 only ) .
p/n 18603-001, Rev. J