CareFusion LTV 1200 Service Manual page 22

Table of Contents

Advertisement

3)
Test each control by pressing every button, one at a time. When each one is pressed,
verify that the name of the button is displayed in the display window.
Control names are as shown below:
Control
Display Select
Breath Rate
Tidal Volume
Pressure Control
Inspiratory Time
Pressure Support
O
% (O
2
Sensitivity
High Pressure Alarm
Low Peak Pressure
Low Minute Volume
Silence Reset
On/Standby
Volume & Pressure
Assist/Control & SIMV/CPAP
Inspiratory / Expiratory Hold
Manual Breath
Low Pressure O
Control Lock
Set Value Knob rotate Left
Set Value Knob rotate Right
PEEP
4)
Test the Set Value knob by turning it clockwise and counterclockwise. Verify that the
direction of rotation is displayed in the display window.
5)
To exit the control test, press Select and the next menu item is displayed.
If the ventilator fails the control test, see Chapter 7 - Troubleshooting for more information.
4
®
Available on the LTV
1200 and 1150 Only.
5
®
Available on the LTV
1200 Only.
p/n 18603-001, Rev. J
4
5
Flush)
2
4
4
5
Source
2
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Display
SELECT
BREATH RATE
TIDAL VOLUME
PRES CONTROL
INSP TIME
PRES SUPPORT
O2%
SENSITIVITY
HIGH PRES
LOW PRES
LOW MIN VOL
SILENCE
ON / STNDBY
MODE VOL/PRS
MODE A/C S/C
IE HOLD
MANUAL BRTH
LOW PRES O2
CONTROL LOCK
ROTATE LEFT
ROTATE RIGHT
PEEP
Page 2-7