Boots, Permanent Removal - CareFusion LTV 1200 Service Manual

Table of Contents

Advertisement

Boots, Permanent Removal

Parts required for replacement:
Replacement Screws Kit, 11149
Replace if damaged:
Finish Washers (6) P/N 10191, or
19119-002
To permanently remove the Upper Protective Boot
1)
Carefully support the disconnected ventilator in an upright position on a clean, dry
surface.
2)
Using a Philips screwdriver, remove the two flat-head mounting screws and finish
washers in the legs of the upper boot and the two flat-head mounting screws and finish
washers in the sides of the upper boot, as shown in the illustration on the next page.
3)
Remove the upper boot and insert and thread two 1/4" pan-head mounting screws of
the correct length into the screw holes in the ventilator's back panel, as shown in the
illustration on the next page.
WARNING
Specific boot screw location – Make sure that the correct screw is being used before
tightening down. Using a screw of the wrong length in this location can cause the ventilator
to malfunction resulting in possible harm to the patient.
31
Contained in CareFusion Replacement Screws kit, P/N 11149.
32
See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for
additional information.
p/n 18603-001, Rev. J
31
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with
torque meter
32
:
Page 8-7