CareFusion LTV 1200 Service Manual page 105

Table of Contents

Advertisement

2)
If the flow output is out of the acceptable output flow range, perform the following, in the
order shown.
 Check all flow valve flexible tubing for leaks or occlusions and replace any
damaged components.
 Recalibrate the flow valve transducer (see Valve Differential Calibration in
Chapter 6 for instructions).
 If flow valve recalibration is indicated, recalibrate the flow valve by adjusting the
VHOME setting (see Flow Valve Calibration in Chapter 6 for instructions).
 If flow valve replacement is indicated, see Flow Valve Assembly in Chapter 8 for
instructions.
O
Blender Checkout
2
Verify the O
blender Solenoid output flows, as follows:
2
1)
Remove the O
blender (see O2 Blender Assembly in Chapter 8 for instructions).
2
2)
Reconnect the 8-wire connector from the O
O
tube Pisco connector to the O
2
3)
Connect a calibrated flow meter to the blender output barb fitting and a 50 psi O
to the blender O
2
4)
Set DIP switch # 5 to ON.
5)
Power the ventilator On (in VENT MTNCE mode), select SOLENOID and set solenoids
# 1 through # 4 to Off.
6)
Verify each (of four) solenoid's output flow is 0.0 Lpm ± 0.5 Lpm.
 If any are not within this range, the O
7)
Select each solenoid individually, set to On and record the output flow.
 If the output flows of all solenoids are within the specified range, the O
correctly calibrated.
 If the output flow of any one of the solenoids is not within the specified range, the O
blender is defective and is to be replaced.
8)
Set DIP switch # 5 to Off, power the ventilator Off and reinstall or replace the O
per instructions O2 Blender Assembly in Chapter 8.
Page 7-28
pressure transducer on the analog board.
2
inlet.
Solenoid
1
2
17.45 to 18.15 Lpm
3
46.36 to 48.24 Lpm
4
46.36 to 48.24 Lpm
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
blender to the power board and the blender
2
blender is defective and is to be replaced.
2
Range
5.79 to 6.01 Lpm
source
2
blender is
2
2
blender
2
p/n 18603-001, Rev. J

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents