Cleaning or replacing the Inlet Filter

To clean the Inlet Filter:
1) Remove the inlet filter by squeezing the foam filter gently with your fingers and pulling it
out.
2) Gently bathe the filter in a solution of mild detergent and warm water.
3) Rinse thoroughly in warm water.
4) Examine the filter for excessive wear or damage. Discard and replace with a new filter if
necessary.
5) Allow the filter to thoroughly air dry before reinstallation.
6) Reinstall the filter.
CAUTION
Wet or Damp Filters - Do not install a wet or damp filter into the LTV
damage the ventilator.
p/n 18603-001, Rev. J
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
®
Ventilator. This could
Page 4-3