General Checkout Worksheet

®
LTV
SERIAL NUMBER:
TEST
DESCRIPTION
Ventilator Checkout ("VENT CHECK")
Alarm
Display
Control
Leak
Vent Inop alarm
Power On Self-Test - External Power
Audible alarm
Display Illumination
Test Messages:
Begin Normal Operation
Volume Operation
Settings:
Volume Mode / Assist/Ctrl
Low Pres O
= Off (LTV
2
BPM = 12
Tidal Volume = 500
Inspiratory Time = 1.0
Pressure Support = - -
®
O
% = 21(LTV
1200 only)
2
Sensitivity = 3
High Pressure Limit = 100
Low Pressure alarm = 5
Low Minute Volume = 1.0
PEEP = 5 cmH
O
2
High PEEP = Off
Low PEEP = Off
High f = Off
p/n 18603-001, Rev. J
____________
STEP
MEAS.
#
VALUE
2)A)
2)B)
2)C)
2)D)
2)E)
3)A)
3)B)
3)C)
3)D)
3)E)
4)A)
®
1200 only)
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
CONDUCTED BY:
DATE:
REQUIREMENT
Audible alarm tone must activate
Confirming audible chirp must activate after
alarm is silenced.
All displays must light except Vent Inop
Correct messages displayed in window
Leak value is less than 1.0 LPM
Audible alarm must sound continuously for
15 seconds and Vent Inop LED must
illuminate continuously for 15 seconds
Confirming audible chirp must activate after
alarm is silenced.
Verify audible alarm sounds for 1 sec
Confirming audible chirp occurs after alarm
All LEDS except Vent Inop illuminate for 3
seconds
Test messages are displayed
Within 20 seconds of power ON
Monitors should read as follows:
Vte:
383 to 633 ml
I:E Ratio:
1:3.8 to 1:4.2
f:
12 bpm
VE:
4.6 to 7.6 L
No alarms are occurring
____________
____________
PASS
(P) /
FAIL
(F)
Page 9-9