ERM
Volume Control Breath
Vte
p/n 18603-001, Rev. J
D
EFINITION
A machine or assist breath where a user-set volume is delivered over a
user-set time. Flow is delivered in a decelerating waveform where the
peak and final flows are calculated so that the final flow is 50% of the
peak flow.
See Tidal Volume, monitored.
®