Table of Contents

Advertisement

®
LTV
1200, 1150,
and 1100 Ventilator
Service Manual
P/N 18603-001 Rev. J

Advertisement

Table of Contents
loading

Summary of Contents for CareFusion LTV 1200

  • Page 1 ® 1200, 1150, and 1100 Ventilator Service Manual P/N 18603-001 Rev. J...
  • Page 2 Revision History Revision Level Revision Date Rev. A, ECO 4188 September 2006 Rev. B, ECO 4950 March 2008 Rev. C, ECO 5407 July 2008 Rev. D, ECO 5640 June 2009 Rev. E, ECO 6002 July 2009 Rev. F, ECO 6291 April 2011 Rev.
  • Page 3 ® 1200, LTV 1150, and LTV 1100 are trademarks belonging to CareFusion Corporation. Copyright © 2006–2013 CareFusion Corporation or one of its subsidiaries. All rights reserved. ® Page ii 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 4  The battery is warranted for ninety (90) days from date of shipment. CareFusion will, at its option, either repair, replace, or issue credit for products that prove to be defective during the warranty period. For warranty service or repair, the product must be returned to CareFusion or a service facility designated by CareFusion, shipping prepaid by the Buyer.
  • Page 5 1200, 1150, or 1100 ventilator, should be directed to a service ® technician who has been trained and certified by CareFusion to service the LTV 1200, 1150, or 1100 ventilator or CareFusion using the contact information at the front of this manual. ® Page iv 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev.
  • Page 6: Table Of Contents

    Contents Notices ............................iv   Chapter 1 - Introduction .................. 1-1   Getting Assistance ........................1-1   Operator’s Safety Information ....................1-2   Warnings ........................... 1-3   Cautions ............................ 1-4   Symbols ............................ 1-5   Chapter 2 - Ventilator Checkout Tests ............2-1  ...
  • Page 7 Solenoid ..........................6-29   Step Test ..........................6-30   Watchdog Test ........................6-31   Configuration ........................... 6-32   Model Selection ........................6-33   Flow Valve Home Position ...................... 6-34   Temperature Compensation ....................6-35   Configuration Menu Exit ......................6-35  ...
  • Page 8 Accumulator ..........................8-53   Alarm Sounder Assembly (previous version) ................. 8-55   Alarm Sounder Assembly (current version) ................8-57   Analog Board Assembly ......................8-59   External Inlet Filter ........................8-63   Fan Assembly ......................... 8-64   Cleaning the Fan Filter ....................... 8-65  ...
  • Page 9 Torque Values ........................E-17   Appendix F - Index ..................F-1   ® Page viii 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 10: Getting Assistance

    LTV 1200, 1150, and 1100 ventilators. It is designed for use by service personnel who have been trained and authorized by CareFusion. Do not perform any of the procedures ® in this manual unless you are trained and authorized for service on LTV 1200, 1150, and 1100 ventilators.
  • Page 11: Operator's Safety Information

    Operator’s Safety Information All Operators must read and understand the following information about Warning, Caution ® and Note statements before operating the LTV 1200, 1150, or 1100 ventilator. General warnings and cautions which apply any time you use the ventilator are listed here. General and specific warnings and cautions also appear throughout this manual where they are most meaningful.
  • Page 12: Warnings

    Unauthorized Parts or Accessories – Serious harm to the patient may result from the use of unauthorized parts or accessories. Only items expressly approved by CareFusion may be ®...
  • Page 13: Cautions

    Cautions CAUTION Electrical Grounding –Loss of protective ground could result in electrical shock. Only use ® the unmodified power supplies and cords supplied for the LTV 1200, 1150, or 1100 ventilator, maintained in good condition and connected to a properly wired and grounded electrical power outlet.
  • Page 14: Symbols

    Symbols Compliance Symbol Title Application ISO 3864 Caution (refer to Direct the user to the instruction (Prev. IEC 348) accompanying manual where it is necessary to follow documents) certain specified instructions where Symbol No. B.3.1 safety is involved. IEC 417 Fuse Indicates the location of fuses.
  • Page 15 ® Page 1-6 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 16 Chapter 2 - V ENTILATOR HECKOUT ESTS This chapter details test procedures that are initiated through the Vent Check menu and ® used to verify the proper operation of the LTV 1200, 1150, or 1100 ventilator. These checkout tests are to be performed before using the ventilator on a patient and in accordance with recommended periodic maintenance and testing of the ventilator.
  • Page 17 To enable the ventilator checkout menu: To enter the ventilator checkout (VENT CHECK) menu, a special power on sequence is required. Disconnect the patient from the ventilator and ventilate using an alternative method. With the ventilator off, connect the AC adapter to the ventilator and plug it in to a valid AC power source.
  • Page 18: Alarm Test

    Alarm Test The alarm test verifies that the audible alarm is working correctly. To run the alarm test: 1) Press Select while ALARM is displayed. 2) Verify that the audible alarm sounds. When the alarm has sounded for at least 2 seconds, press Select again.
  • Page 19: Display Test

    Display Test At the end of the alarm test, DISPLAY shows in the LED display window. The display test verifies that the ventilator displays are working correctly. To run the display test: 1) Press Select while DISPLAY is showing. 2) All segments of the 7-segment control displays; all dots of the dot-matrix window displays and all LEDs are illuminated.
  • Page 20 Displays illuminate in the following colors: Display Color Display Color Airway Pressure Display Green Assist/Control Mode LED Green Display Window SIMV/CPAP Mode LED Green Breath Rate Green NPPV Mode LED Green Tidal Volume Green Inspiratory / Expiratory Hold Green Pressure Control Green Manual Breath LED Green...
  • Page 21: Control Test

    Control Test At the end of the display test, CONTROL shows in the LED display window. The control test is used to verify that the ventilator buttons and the Set Value knob are working correctly. To run the control test: Press Select while CONTROL is displayed.
  • Page 22 Test each control by pressing every button, one at a time. When each one is pressed, verify that the name of the button is displayed in the display window. Control names are as shown below: Control Display Display Select SELECT Breath Rate BREATH RATE Tidal Volume...
  • Page 23: Leak Test

    Leak Test The leak test is used to test the patient circuit for leaks. To run the leak test: 1) At the end of the control test, LEAK shows in the LED display window. 2) Attach a patient circuit with all accessories (such as water traps, heated circuits and humidifiers to the patient circuit).
  • Page 24 6) To perform the leak test, the ventilator does the following: a) Closes the exhalation valve and sets the flow valve to a near-closed state. The display briefly shows HOMING VALVE. b) Elevates the turbine motor speed. The display shows SET TURBINE. c) Elevates the circuit pressure.
  • Page 25: Vent Inop Alarm Test

    Vent Inop Alarm Test The Vent Inop alarm test is used to verify that the Vent Inop alarm is working correctly. To run the Vent Inop alarm test: To run the Vent Inop alarm test, the ventilator must be on (running) for at least 60 seconds.
  • Page 26: Exit

    Exit To exit VENT CHECK and enter normal ventilation mode: 1) Turn the Set Value knob to scroll through the main menu entries (VENT OP, ALARM OP, VENT CHECK, etc.) until EXIT is displayed. 2) Press Select while EXIT is displayed. Alternatively, press Control Lock repeatedly until normal ventilation mode is restored.
  • Page 27 ® Page 2-12 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 28 Chapter 3 - R RANSDUCER The real-time transducer data allows you to view the real-time activity in the ventilator. The menu is set up as follows: RT XDCR DATA xx.xx xx.xx xx.xx xx.xx FTw or FTn xx.xx x.xx LEAK xx.xx xx.xx xx.xx STEP...
  • Page 29 Display Real-time Data FDn xx.xx Flow differential pressure as measured at the patient wye using the narrow scale transducer. Differential pressure is measured between the high and low side proximal sense lines. The narrow scale transducer is only used for differential pressures between -0.35 cmH O and 0.35 cmH O (approximately -15...
  • Page 30: Cleaning The Ventilator

    Chapter 4 - C LEANING ISINFECTING AND TERILIZING Cleaning the Ventilator All ventilator external surfaces should be cleaned before and after each patient use, and as otherwise necessary. To clean the ventilator: 1) Wipe the exterior surfaces of the ventilator with a clean, damp cloth. The use of an anti- bacterial cleaning solution is recommended.
  • Page 31: Cleaning Or Replacing The Fan Filter

    Cleaning or replacing the Fan Filter To clean the fan filter: Using a small screwdriver, detach the fan filter grill from its housing. Remove the fan filter by squeezing the foam filter gently with your fingers and pulling it out. NOTE Hardware Fault -If you touch the fan blades while removing the fan filter grill or filter, a HW FAULT may occur.
  • Page 32: Cleaning Or Replacing The Inlet Filter

    Cleaning or replacing the Inlet Filter To clean the Inlet Filter: 1) Remove the inlet filter by squeezing the foam filter gently with your fingers and pulling it out. 2) Gently bathe the filter in a solution of mild detergent and warm water. 3) Rinse thoroughly in warm water.
  • Page 33: Cleaning Or Replacing The O Inlet Filter (Ltv 1200 Only)

    O Inlet Filter (P/N 14313 or P/N 19845- 001) and O-Ring (P/N 10609), available from CareFusion. In addition to the existing internal O Inlet filter, P/N 14313 or P/N 19845-001, an External, In-Line Oxygen Filter (P/N 14470) is available from CareFusion.
  • Page 34 5) Replace the filter by sliding it back into the O inlet port. Replace the O-Ring, making sure it is completely tucked under the retaining lip on the inside of the O inlet port. 6) Reconnect the high pressure O line.
  • Page 35: Cleaning The Reusable Exhalation Valve And Patient Circuit

    Cleaning the Reusable Exhalation Valve and Patient Circuit WARNING Patient Circuits – CareFusion Patient Circuits, Exhalation Valve Assemblies and Water Traps are shipped clean, not sterile. ® ® ® Refer to your LTV 1200, LTV 1150, or LTV 1100 Ventilator Operator’s Manual or the Instructions for Use enclosed with your patient circuits and patient circuit components for assembly, disassembly, care and cleaning recomendations.
  • Page 36: Recommended Maintenance Schedule

    10,000 hour, two year and/or 30,000 hour, six year extended maintenance and ventilator repair must be performed by a service technician who has been trained and certified by CareFusion. Replacement at 10,000 hours or 2 years is based on normal use of up to 200 charge cycles. The battery may need to be replaced more frequently if it is being charged more often.
  • Page 37: Before Initial Use

    Before Initial Use Plug the ventilator into an AC power source for 24 hours to fully charge the internal battery. While charging in the standby mode, the Charge Status Indicator LED displays as flashing amber, solid amber or solid green. If the Charge Status Indicator displays red, the internal battery cannot be charged and should be replaced.
  • Page 38: 10,000 Hour / 2 Year Maintenance

    1200, 1150, and 1100. Do not attempt to service the ventilator unless you have been properly trained to do so, personal injury could result. Refer all servicing and repair to CareFusion or an authorized service center. 10,000 Hour / 2 Year Maintenance...
  • Page 39 ® Page 5-4 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 40: Chapter 6 - Maintenance & Calibration

    Chapter 6 - M & C AINTENANCE ALIBRATION Operating Theory ® The LTV 1200, 1150, and 1100 ventilator utilizes an electromechanical pneumatic system under the control of a microprocessor to deliver patient ventilation. The following diagram and descriptions illustrate the major components of the ventilator and their respective functions.
  • Page 41 Ventilation gas exiting the Flow Valve is connected to the Wye and Exhalation Valve by a patient circuit. The Flow Sensor at the Wye measures the exhaled flow using a fixed orifice type transducer. Transducer sensor ports are located between the patient and ventilator connection ports. The Exhalation Valve provides the following functions: 1) Closes the exhalation port during inspiration to divert gas to the patient.
  • Page 42: Pneumatic Schematic Detail

    Pneumatic Schematic Detail ® p/n 18603-001, Rev. J 1200, 1150, and 1100 Ventilator Service Manual Page 6-3...
  • Page 43: Vent Maintenance

    Vent Maintenance The ventilator maintenance tests are used to verify the ventilator is working correctly, troubleshoot problems with performance, and perform maintenance procedures such as calibration. The maintenance menu is set up as follows: VENT MTNCE CALIBRATION SERVO SOLENOID STEP TEST WDOG TEST CONFIG CLEAR...
  • Page 44: Entering Vent Maintenance Mode

    Entering Vent Maintenance Mode WARNING ® Ventilator Checkout and Maintenance Modes - The LTV 1200, 1150, and 1100 ventilator does not deliver gas during the ventilator Checkout mode (VENT CHECK) or ventilator Maintenance mode (VENT MTNCE) and should not be used to ventilate a patient during these tests.
  • Page 45: Vent Maintenance Entry Alarm

    Vent Maintenance Entry Alarm When you power the ventilator on in Vent Maintenance mode, REMOVE PTNT is displayed and the audible alarm sounds to remind you to remove the patient from the ventilator and use an alternate method of ventilation. To clear the REMOVE PTNT alarm: 1) Press the Silence Reset button.
  • Page 46 1100 only - The LTV 1100 Ventilator shall be calibrated while in the Flow Sensing Patient Circuit Mode (FL SENSOR). The patient circuit mode Default Setting is Flow Sensor. Contact CareFusion or refer to Chapter 12 – Operating Procedure in ® the LTV 1100 Operators Manual P/N 21106-001 for detailed information.
  • Page 47: Airway Pressure Calibration

    Airway Pressure Calibration To perform the airway pressure calibration, you will need the following equipment:  Calibration Syringe assembly , P/N 11471, with a T-connection to a pressure manometer (0-90 cmH The Calibration Syringe is available separately, or as part of the Maintenance Calibration Kit, P/N 11566. A 60 cc/mL Syringe with a calibration U-Tube test hose (P/N 18559-001) may also be used.
  • Page 48 11) If leakage in excess of the required values is observed, troubleshoot leaks (see Chapter 7 - Troubleshooting). If the leak persists, contact the Technical Support department at CareFusion using the information at the front of this manual. ® p/n 18603-001, Rev. J...
  • Page 49: Calibrate The Airway Pressure Transducers

    Calibrate the Airway Pressure Transducers Press the Select button while AIRWAY is displayed. AP 50 O is displayed. If necessary, disconnect the High and Low pressure sense lines from the side of the ventilator: Connect the male Luer fitting from the calibration syringe assembly to the high pressure female Luer fitting on the ventilator.
  • Page 50: Flow Differential Calibration

    Flow Differential Calibration To perform the flow differential calibration, you will need the following equipment:  Calibration syringe assembly , P/N 11471, with a T-connection to a pressure manometer (0-90 cmH O). The pinch clamps should be used between each Luer fitting and the T-connection to block off the unused Luer fitting for each step in this procedure.
  • Page 51 2) Connect the female Luer fitting from the calibration syringe assembly to the low pressure male Luer fitting. Pinch off the tubing on the calibration syringe assembly to the male Luer fitting, and increase and maintain the pressure at 30.0, ± 0.2 cmH Observe the numbers displayed in the Tidal Volume setting window.
  • Page 52 6) Connect the male Luer fitting from the calibration syringe assembly to the high pressure female Luer fitting on the side of the ventilator. Decrease and maintain the vacuum until a value in the range of 10 to 150 is displayed in the Tidal Volume window. If necessary, add a large amount of compliance, such as a test lung, to the calibration syringe assembly, this will significantly stabilize the readings during this step.
  • Page 53: Valve Differential Calibration

    Valve Differential Calibration To perform the valve differential calibration, you will need the following equipment:  Calibration syringe assembly , P/N 11471, with a T-connection to a pressure manometer (0-90 cmH The Calibration Syringe is available separately, or as part of the Maintenance Calibration Kit, P/N 11566. A 60 cc/mL Syringe with a calibration U-Tube test hose (P/N 18559-001) may also be used.
  • Page 54 To calibrate the valve differential transducer: Press the Select button while VALVE DIFF is displayed. VD 15 O is displayed. Disconnect both flexible tubes from the valve differential transducer on the analog board. NOTE See instructions beginning on page 8-33 for tube routing diagrams, tables and instructions. ®...
  • Page 55 Remove a Luer fitting from the calibration syringe assembly. Connect the tube from the syringe to the high pressure port of the valve differential transducer (the port nearest the analog board). Use the pinch clamp on the unused line from the calibration syringe assembly to hold the pressure steady.
  • Page 56: Peep Pilot Pressure Transducer Calibration

    PEEP Pilot Pressure Transducer Calibration To calibrate the PEEP Pilot Transducer: 1) Press the Select button while PPP is displayed. PPP 30 O is displayed. 2) Remove the tube connected to the PEEP pilot pressure transducer (PT2 on the analog PCBA).
  • Page 57: O 2 Inlet Pressure Calibration (Ltv ® 1200 Only)

    Inlet Pressure Calibration (LTV 1200 Only) ® To perform the O inlet pressure calibration, you will need the following equipment:  Gas source capable of 80 PSI with a T-connection to a pressure gauge and an O inlet connector. leak test Remove the O inlet port cap and connect the high pressure gas source to the O inlet...
  • Page 58 Check the values displayed in the Tidal Volume window. When the value stabilizes, record the number on the Calibration Worksheet and press Select. O2 0 PSI is displayed. With the O inlet port open to ambient room air, observe the values in the Tidal Volume window.
  • Page 59: Stepper Motor Calibration

    Stepper Motor Calibration To perform the stepper motor calibration, you will need the following equipment:  Stepper Motor Calibration Tool , P/N 10871  10 amp 60 Hz amp meter set to measure between 400 and 700 ma. To calibrate the motor drive: CAUTION Possible ventilator damage - To avoid damaging the ventilator, turn the ventilator off and disconnect the AC adapter before attaching the calibration tool.
  • Page 60 Set the calibration tool to ΦA. Reconnect the AC adapter to the ventilator. Turn the ventilator on, silence alarms and enter the CALIBRATION menu. Turn to MOTOR DRIVE and press Select. ΦA:+600ma: xxx is displayed where the xxx is a numeric value. The Phase A amp meter will show a positive value.
  • Page 61: Measure Actual Turbine Speed (T/S)

    Measure Actual Turbine Speed (T/S) Enter the SERVO menu. Run the ventilator in Servo mode at default settings (flow 10 and turbine speed 4000) for a minimum of one minute to stabilize turbine speed. Set up a multimeter to measure frequency. Connect the meter test leads as follows: ...
  • Page 62 Performance Checkout patient port pressure testing. If the actual T/S value is outside of the above range, check the setup and repeat the test. For assistance, call Technical Support at CareFusion using the contact information at the front of this manual.
  • Page 63: Vent Maintenance Exit

    Vent Maintenance Exit The ventilator will not resume normal operation while the maintenance mode DIP is set. To exit maintenance mode: 1) Turn the ventilator off. 2) Locate DIP #5 and set it to OFF (towards the outside of the ventilator). Verify that DIP switches #6 and #8 remain ON.
  • Page 64: Calibration Worksheet

    Enter displayed value For Software Version < 05.04 1915 - 2340 For Software Version ≥ 05.04 1915 - 2521 VD @ Ambient (LTV 1200/1150) Enter displayed value 40 - 328 VD @ Ambient (LTV 1100) Enter displayed Value 61 - 354...
  • Page 65: Flow Valve Calibration (Vhome Trim)

    Flow Valve Calibration (VHome Trim) The flow valve should not normally require recalibration from its factory set values. However, some circumstances, such as mechanical shock, may cause the flow valve calibration to change from its factory-set condition. This flow valve calibration procedure can be used to adjust the flow valve calibration setting to compensate for improper flows measured during the Performance Checkout procedure.
  • Page 66 Read the flow value from the flow meter. Correct flow shall be 9.4 to 9.8 Lpm (ATP) or 9.7 to 10.0 Lpm (BTPD).  If the flow is between 9.4 to 9.8 Lpm (ATP) or 9.7 to 10.0 Lpm (BTPD), no flow vlave calibration is needed ...
  • Page 67: Servo

    Servo The Servo feature allows you to verify certain ventilator functions. When SERVO is on, the ventilator continuously delivers gas according to the highlighted settings. Gas may be delivered by:  turbine speed and flow valve step position  turbine speed and flow in Lpm SERVO is used to check ventilator operation or diagnose problems by setting specific conditions on the ventilator and monitoring the system either externally (e.g.
  • Page 68: Solenoid

    Solenoid The Solenoid menu is used to test the operation of the solenoids. The menu is set up as follows: SOLENOID ExhPilot OFF Purge OFF APres OFF ExhDiffP OFF PPin OFF PPout OFF ® O2 #1 OFF (LTV 1200 only) ®...
  • Page 69: Step Test

    Step Test The flow valve stepper motor synchronization test is used to exercise the flow valve and ® insure LTV software / hardware synchronization. 1) Set the ventilator to VENT MAINTENANCE mode (VENT MTNCE), (refer to Vent Maintenance Entry in this chapter). 2) From the VENT MTNCE display, press Select and rotate the Set Value knob to select the STEP TEST display.
  • Page 70: Watchdog Test

    Watchdog Test The watchdog timer is used to verify that essential parts of the software are running at the correct times. If the software does not update the watchdog timer correctly, the watchdog timer causes the ventilator to inop. This item tests that the watchdog timer is operating correctly.
  • Page 71: Configuration

    The menu selections are described further in the following pages. NOTE ® The configuration menus are used to configure many models of LTV The Serial Number menu option is for use by CareFusion personnel only. ® Page 6-32 1200, 1150, and 1100 Ventilator Service Manual...
  • Page 72: Model Selection

    The model selected should match the model number on the front of the ventilator. The menu is set up as follows: MODEL LTV 1200 LTV 1150 LTV 1100 LTV 1000...
  • Page 73: Flow Valve Home Position

    Flow Valve Home Position The flow valve home position menu is used to select the home step position for the flow valve. The VHome value is printed on the VHome label of the flow valve. The menu is set up as follows: FLOW VALVE Range...
  • Page 74: Temperature Compensation

    Temperature Compensation The Temperature Compensation menu is used to set the temperature compensation for the flow valve ON or OFF. NOTE ® ® This configuration menu is used for many models of LTV . For the LTV 1200, 1150, and 1100, which always has a Thermistor installed, the setting for Temperature Compensation is always ON.
  • Page 75: Clear

    CLEAR The CLEAR menu is used to clear sections of the non-volatile EEPROM memory. The menu is set up as follows: CLEAR CLR EVENTS CLR CONTROLS CLR CAL CLR BAD REC CLR ALL CLR EXIT Use the Set Value knob to rotate between CLEAR options. To clear data from a section of the EEPROM, press Select while the appropriate name is displayed in the window.
  • Page 76: Vent Maintenance Menu Exit

    Vent Maintenance Menu Exit To exit the vent maintenance menu: 1) Press Select while EXIT is displayed. 2) VENT MTNCE is displayed. To access the other main Extended Features menu options, turn the Set Value knob to the desired item and press Select. 3) To return the ventilator to normal operation, see the instructions under Vent Maintenance Exit at the beginning of this section.
  • Page 77 ® Page 6-38 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 78: Chapter 7 - Troubleshooting

    ® to perform repairs or maintenance on LTV 1200, 1150, and 1100. Do not attempt to repair or replace any part of the ventilator unless you are trained and certified by CareFusion. Personal injury could result. NOTE Event log entries are only one of many diagnostic tools used to troubleshoot the ventilator.
  • Page 79: Displays And Buttons

    You may also be able to diagnose the problem by reading sections with related symptoms. For information on resolving problems that are not listed here, contact CareFusion using the contact information at the front of this manual.
  • Page 80 Symptoms Possible Causes What to Do Low Min Vol display LOW MIN VOL alarm The Low Min Vol display is flashed and flashing. occurred. the LOW MIN VOL message is displayed when a low minute volume alarm occurs. The display will continue to flash even ®...
  • Page 81 Symptoms Possible Causes What to Do A control doesn't Control not active in If a control is dimmed, it is not active in the activate. selected mode. currently selected mode and changing its setting does not affect ventilation. (See the Set Value knob ®...
  • Page 82 Symptoms Possible Causes What to Do Rotary switch is Verify the rotary switch is properly disconnected. connected. If necessary, replace the rotary switch assembly. Defective rotary switch. See Rotary Switch (Set Value Knob) Assembly in Chapter 8 for instructions. Defective main board. Replace the main board.
  • Page 83 Symptoms Possible Causes What to Do LO PEEP OFF is The Low PEEP alarm is This is an informational message only ® displayed. turned off. (See the LTV 1200, 1150, or 1100 Ventilator Operator's Manual for an explanation of this feature). SBT f OFF is The SBT High Breath Rate This is an informational message only...
  • Page 84: Ventilator Performance

    Ventilator Performance Symptoms Possible Causes What to Do Ventilator is auto Sense lines are reversed. The sense lines are not designed to be cycling, monitored removed from either the wye or the Luer volumes are very fittings. If the sense lines have been small, and RT XDCR removed and replaced incorrectly, they DATA item FTx...
  • Page 85 Symptoms Possible Causes What to Do Monitored volume is Very small ET tube A very small ET tube connected directly to high. connected directly to wye. the wye may cause turbulence that causes the flow differential to be read incorrectly. Delivered volume is To reduce this turbulence, add a short high.
  • Page 86 Symptoms Possible Causes What to Do CONTINUED… Defective flow valve. Calibrate the flow valve. See Flow Valve Calibration in Chapter 6 for instructions. Monitored volume is high. Clean the Flow Valve following the Flow Valve Cleaning Procedure, see page 8-66. Delivered volume is Perform the Advanced Vte Diagnostic high.
  • Page 87 Symptoms Possible Causes What to Do CONTINUED … Leak Compensation is not Verify that the Leak Compensation Extended Features option is set to On Monitored volume is ® (default setting is on). See the LTV 1200, low. 1150, or 1100 Ventilator Operator's Delivered volume is Manual for more information.
  • Page 88 Symptoms Possible Causes What to Do Delivered pressure is Circuit leak. Run a Leak Test and reseat or replace the low, PEEP is low, leaking parts or connections. See Leak ventilator is auto Test in Chapter 6 for instructions. cycling. High or low side sense line Check high and low pressure sense lines Delivered pressure is...
  • Page 89 Symptoms Possible Causes What to Do CONTINUED … Failed auto zero. Perform an auto zero under XDCR ZERO. ® See the LTV 1200, 1150, or 1100 Delivered pressure is Ventilator Operator's Manual for more low, PEEP is low, information. ventilator is auto cycling.
  • Page 90 Symptoms Possible Causes What to Do Delivered pressure is Diaphragm is incorrectly Open the exhalation valve and remove the high. seated in exhalation valve. diaphragm. Reseat the diaphragm and snap the valve cap back in place. See Monitored pressure is Cleaning the Reusable Exhalation Valve high.
  • Page 91 Symptoms Possible Causes What to Do Delivered flow is high. High or low side sense line Check high and low pressure sense lines or elbow at patient wye to be sure they are correctly attached and Delivered flow is low. loose or leaking.
  • Page 92 Symptoms Possible Causes What to Do Sensitivity does not Circuit leak. Run a Leak Test and reseat or replace the appear to be accurate. leaking parts or connections. See Leak Test in Chapter 2 for instructions. Ventilator is auto cycling. Sense lines are reversed.
  • Page 93 Source correctly calculated and set using the selected. Input O Flow Chart. inlet flow too high when CareFusion recommends the use of an O Low O Source selected. monitor to verify delivered O %. Adjust the entrained O flow so the monitored value ®...
  • Page 94 Low O Source selected. correctly calculated and set using the Input O Flow Chart. CareFusion recommends the use of an O monitor to verify delivered O %. Adjust the entrained O flow so the monitored value ®...
  • Page 95 Symptoms Possible Causes What to Do PEEP not working. Circuit leak. Run a Leak Test and reseat or replace the leaking parts or connections. See Leak PEEP low. Test in Chapter 2 for instructions. PEEP sags during exhalation. Diaphragm incorrectly Open the exhalation valve and remove the seated in exhalation valve.
  • Page 96 Symptoms Possible Causes What to Do Internal flexible tubing CONTINUED … Open the vent and verify that none of the occluded or pinched. flexible tubes connected to the PEEP not working. accumulator, solenoid mount assembly, Internal flexible tubing has PEEP low. solenoid manifold, analog board, oxygen pinhole leaks or leaking at PEEP sags during...
  • Page 97 Symptoms Possible Causes What to Do Verify the sensitivity is set to an Ventilator is auto Circuit leak. appropriate value. For a high leak cycling environment, the sensitivity may need to Sensitivity does not be set higher to prevent auto cycling. appear to be accurate.
  • Page 98 Symptoms Possible Causes What to Do High or low sense lines are Verify lines are not occluded or pinched Condensation in sense occluded. and/or clear the lines with a low flow (less lines. than 10 Lpm) gas source. High or low sense ports in the wye are occluded.
  • Page 99 Symptoms Possible Causes What to Do When the battery charge circuit is running Ventilator makes a Battery charge circuit in bulk charge (the Charge Status LED is high pitched noise running. amber) the ventilator may emit a high when in Standby. pitched sound that some people can hear.
  • Page 100: Advanced Diagnostic Procedures

    Advanced Diagnostic Procedures Advanced Vte Diagnostic Procedures If the ventilator Monitored Exhaled Volume (Vte) readout, Delivered Volume to the lung, Target Pressures or Inspired Volumes (LTM only) appear incorrect; perform the following checkout procedures.  When instructed to replace a subassembly/component that fails any of the following checkout procedures, verify the original problem is resolved once the subassembly/component has been replaced.
  • Page 101 Allow the unit to run for 1 minute and record the displayed Vte (average for 8 consecutive breaths). Then stand the unit upright, allow it to run for 1 minute and record the displayed Vte (average for 8 consecutive breaths). ...
  • Page 102 FVd Delivered Volume: Apply 15 cmH O to the flow valve differential transducer at the high side port and verify the FVd readout is within ± 0.5 cmH O of the applied pressure. Disconnect the tube from the syringe to the high pressure port of the valve differential transducer so the connection is open to ambient air.
  • Page 103 Flow Valve Accuracy Checkout Check flow valve accuracy as follows: At the Tidal Volume setting which is giving failing results, press the Tidal Volume button and VCALC xxx Lpm will be displayed (this is the calculated peak flow for the set volume).
  • Page 104: Advanced Fio 2 Diagnostic Procedures (Ltv ® 1200 Only)

    Advanced FiO Diagnostic Procedures (LTV 1200 Only) ® Delivery of the correct percentage of oxygen (FiO ) is dependent on correct calibration of the oxygen pressure transducer, flow valve and O blender. If the FiO appears incorrect, perform the following diagnostic procedures. When instructed to replace a subassembly or component that fails any of the following checks, verify the original problem is resolved once the item has been replaced.
  • Page 105 If the flow output is out of the acceptable output flow range, perform the following, in the order shown.  Check all flow valve flexible tubing for leaks or occlusions and replace any damaged components.  Recalibrate the flow valve transducer (see Valve Differential Calibration in Chapter 6 for instructions).
  • Page 106: Power And Battery Operation

    Power and Battery Operation Problem Possible Causes What To Do The ventilator does not Defective AC source or Connect the ventilator to a known good AC power up. adapter and depleted source using a known good AC adapter. internal battery. Verify the power cord for the adapter is fully seated.
  • Page 107 Problem Possible Causes What To Do The ventilator doesn’t Defective AC source. Make sure the AC adapter is plugged into operate from external a good AC source and is securely AC adapter power cord power. connected to the ventilator. Verify the loose.
  • Page 108 Problem Possible Causes What To Do Battery Charge Status Internal battery charging. The Charge Status LED flashes amber LED is flashing amber. while the battery charging circuit evaluates the battery as a part of the charge cycle. If the battery is found to be OK, the Charge Status LED will change to solid amber while the battery is charging.
  • Page 109: Alarms

    Operator's Manual. Single occurrences of some alarms, such as HW FAULT or RESET may be caused by ESD. If these alarms recur, and for other alarms that do not usually occur during normal operation, follow the instructions in this section or contact CareFusion using the contact information at the front of this manual.
  • Page 110 Symptoms Possible Causes What to Do Ventilator won’t Diaphragm installed Open the exhalation valve and remove the exhale, repeated backwards or incorrectly diaphragm. Reseat the diaphragm and HIGH PRES alarms, seated in exhalation valve. snap the valve cap back in place. See turbine stops and Cleaning the Reusable Exhalation Valve pressure drops, then...
  • Page 111 Symptoms Possible Causes What to Do CONTINUED … Exhalation drive line leaking Check the exhalation drive line at both the or loose. ventilator and exhalation valve ends. Verify Repeated the line is securely seated and not leaking. DISC/SENSE alarms. Exhalation valve leaking during inspiration.
  • Page 112 Symptoms Possible Causes What to Do HW FAULT alarm. HW FAULT alarms: Check the Event Trace to determine the kind of error and follow the instructions AD MMTCH below for the specific error type. (See the AD MTCH1 ® 1200, 1150, or 1100 Ventilator EEPROM Operator's Manual for instructions on using FAN FLT1...
  • Page 113 Board in Chapter 8 for instructions. Defective main board. Replace the main board. See Main Board Assembly in Chapter 8 for instructions. All other causes. Contact CareFusion. NO CAL DATA alarm. Failed or missing calibration Recalibrate the vent. See Chapter 6 - records.
  • Page 114 Operator's Manual for instructions). Defective Remote Alarm Contact Remote Alarm System System. manufacturer or service personnel. Internal problem with the Immediately contact a certified CareFusion ventilator. service technician. Remote Alarm PNT ASSIST option set to Set PNT ASSIST option to NORMAL. (See ®...
  • Page 115 Defective Remote Alarm Contact Remote Alarm System System. manufacturer or service personnel. Internal problem with the Immediately contact a certified CareFusion ventilator. service technician. Patient Assist Call Incorrect Patient Assist Establish whether the Patient Assist Call System does not work...
  • Page 116 What to Do CONTINUED … Internal problem with the If problem reoccurs, immediately contact a ventilator. certified CareFusion service technician. Repeated SBT < f alarms. Repeated SBT > f/Vt Total Breath Rate (f) Check SBT > f/Vt alarm value. (See the ®...
  • Page 117: Checkout Test Failures

    Checkout Test Failures Symptoms Possible Causes What to Do ALARM Test Alarm volume set too high. Set the alarm volume under the Extended ® Features Menu. (See the LTV 1200, Audible alarm too loud. 1150, or 1100 Ventilator Operator's Manual for an explanation of the ALARM VOL feature.) ALARM Test...
  • Page 118 Symptoms Possible Causes What to Do Control Test Wrong model selected in Verify the correct MODEL has been maintenance mode. selected in VENT MTNCE. See Chapter 6 No message is - Maintenance & Calibration. displayed when some control buttons are pressed.
  • Page 119 LED did not last 15 seconds (fail). Discontinue use of the ventilator and Vent Inop Alarm Test Unknown contact CareFusion for information. Vent Inop alarm tone and/or LED fails test. Repeat the Vent Inop Alarm Test and allow Vent Inop Alarm Test...
  • Page 120: Test Lung Operation

    Pressure> 40 cmH O used The compliance characteristics of some higher than set on small test lung small test lungs (CareFusion or Siemens pressure on test lung. (CareFusion or Siemens 190) cause incorrect readings when high 190.) pressures are used. For these lungs, use...
  • Page 121 ® Page 7-44 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 122: Chapter 8 - Component Removal And Replacement

    Training and Authorization This manual is only intended for use by service personnel who have been trained and authorized by CareFusion. Do not perform any of the procedures in this manual unless you ® are trained and authorized for service on LTV 1200, 1150, and 1100 ventilators.
  • Page 123: Tools

    CareFusion Respiratory Systems 22745 Savi Ranch Parkway Yorba Linda, California 92887-4645, USA Customer Care Phone: 800.754.1914 763.398.8500 Customer Care Fax: 763.398.8403 Customer Care Center E-mail: ltvservice@carefusion.com Website: www.carefusion.com ® Page 8-2 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 124: Before Removing The Back Panel

    Before removing the back panel CAUTION Opening the ventilator - Always turn the ventilator OFF and remove the external power before opening the ventilator case or attempting to service the ventilator. Anti-static precautions - Always wear a grounded anti-static wrist strap when handling the ventilator with the case open.
  • Page 125: Boots, Protective

    Boots, Protective Rubberized protective boots are installed on the top and bottom of all current versions of ® ventilators to protect against accidental shocks and strikes to the casing. This section is divided into four sub-sections to accommodate the removal, replacement, or installation of the boots under the following conditions: ...
  • Page 126: Boots, Temporary Removal

    Retain the screws and finish washers for reuse when the upper boot is reinstalled. Remove the upper boot. Contained in CareFusion Replacement Screws kit, P/N 11149. See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for additional information.
  • Page 127 To temporarily remove the Lower Protective Boot Lay the ventilator down (front up) and use a Philips-head screwdriver to remove the four flat-head mounting screws and finish washers in the sides of the lower boot, as indicated in the illustration below. Retain the screws and finish washers for reuse when the lower boot is reinstalled.
  • Page 128: Boots, Permanent Removal

    Using a screw of the wrong length in this location can cause the ventilator to malfunction resulting in possible harm to the patient. Contained in CareFusion Replacement Screws kit, P/N 11149. See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for additional information.
  • Page 129 Insert and thread two 1/4” flat-head mounting screws with finish washers into the screw holes in the ventilator’s side panels, as shown in the illustration.  Finish washers should be already in place. Torque tighten the mounting screws to these specified values ...
  • Page 130 To permanently remove the lower protective boot Lay the ventilator down (front up) and use a Philips-head screwdriver to remove the four flat-head mounting screws and finish washers in the sides of the Lower Protective Boot, as indicated in the illustration below. Remove the lower boot and insert and thread four 1/4”...
  • Page 131: Boots, Reinstallation

    Specific boot screw location –Using a screw of the wrong length in this location can cause the ventilator to malfunction resulting in possible harm to the patient. Contained in CareFusion Replacement Screws Kit, P/N 11149 See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for additional information.
  • Page 132 Torque tighten the mounting screws to these specified values.  Torque tighten the screws in the legs of the boot to 60 in-oz (0.42 Nm)  Torque tighten the screws in the sides of the boot to 20 in-oz (0.14 Nm) CAUTION Damage to finish washers -Do not over tighten screws to avoid damage to the finish washers.
  • Page 133 To reinstall the lower protective boot Lay the ventilator down (front up) and orient the lower boot to the ventilator as shown in the illustration below. Move the boot into position on the bottom of the ventilator and align its four screw holes with the corresponding holes in the ventilator side panels.
  • Page 134: Boots, New Installation

    Using a screw of the wrong length in this location can cause the ventilator to malfunction resulting in possible harm to the patient. Contained in CareFusion Protective Boots Installation Kit, P/N 11550 Contained in CareFusion Replacement Screws kit, P/N 11149.
  • Page 135 Torque tighten the mounting screws to these specified values  Torque tighten the screws in the legs of the boot to 60 in-oz (0.42 Nm)  Torque tighten the screws in the sides of the boot to 20 in-oz (0.14 Nm) CAUTION Damage to finish washers -Do not over tighten screws to avoid damage to the finish washers.
  • Page 136 To install the lower protective boot Lay the ventilator down (front up) and use a Philips-head screwdriver to remove the four flat-head mounting screws in the ventilator’s side panels, as indicated in the illustration below.  Do not remove the mating finish washers. Orient the lower boot to the ventilator as shown in the illustration below.
  • Page 137: Ltm/Ltv Mounting Assembly Installation And Removal

    LTM/LTV mounting assembly installation and removal Parts Required for Installation: Tools required: Components  Phillips screwdriver ® The current LTM™ to LTV mounting assembly, p/n with torque meter 17916-001, which contains:  7/64” Allen wrench  ® LTM™ to LTV mounting bracket (1), p/n 17918-001 ®...
  • Page 138: Current Model Ltm Instructions

    Current Model LTM Instructions Temporary removal (current model LTM) Parts required for replacement: Tools required:   None Phillips screwdriver with torque meter Use a small Philips screwdriver and remove the five back panel mounting screws and the LTM mounting assembly, as shown. Retain the screws for reuse when the mounting assembly is reinstalled.
  • Page 139: Ltm Mounting Assembly Permanent Removal (Current Ltm)

    LTM mounting assembly permanent removal (current LTM) Parts required for replacement: Tools required:   Replacement Screws Kit, P/N Phillips screwdriver with torque 11149 meter ® Remove the LTM/ LTV mounting assembly (see instructions on page 8-17). Insert five 1/4” back panel replacement mounting screws (from the replacement screws kit, P/N 11149) and torque tighten to 60 in-oz (0.42 Nm).
  • Page 140: Ltm Mounting Assembly Re-Installation (Current Ltm)

    LTM mounting assembly re-installation (current LTM) Parts required for replacement: Tools required:   Replacement Screws Kit, P/N 11149 Phillips screwdriver with torque meter Position the ventilator and mounting assembly as shown in the illustration below. Place the mounting assembly onto the ventilator and align the screw slots with the corresponding screw holes in the ventilator.
  • Page 141: Ltm Mounting Assembly New Installation (Current Ltm)

    LTM mounting assembly new installation (current LTM) Remove the upper boot (if installed). See Boots, Protective in this chapter for instructions. WARNING Mounting screw use – Internal damage to the ventilator may result if the wrong length mounting screws are used when installing or removing external accessories. Accessory mounting screws - Refer to the information supplied with replacement screws kit, P/N 11149, to determine the appropriate accessory mounting screws or accessory ®...
  • Page 142 To attach the LTM mounting assembly (p/n 17916-001): 1) Place the disconnected ventilator on a clean, dry surface. 2) Use a Philips screwdriver and remove the five screws from the top and back panel of the ventilator, as indicated in the illustration.
  • Page 143: Ltm/Ltv ® Mounting Bracket (Earlier Model Ltm)

    LTM/LTV mounting bracket (earlier model LTM) ® Earlier model LTM Graphics Monitors require a two piece mounting assembly as shown below. WARNING Mounting screw use – Internal damage to the ventilator may result if the wrong length mounting screws are used when installing or removing external accessories. Accessory mounting screws - Refer to the information supplied with the replacement screws kit, P/N 11149, to determine the appropriate accessory mounting screws or accessory replacement screws to use when removing or exchanging external accessories...
  • Page 144: Ltm/Ltv Mounting Bracket Temporary Removal (Earlier Model Ltm)

    LTM/LTV mounting bracket temporary removal (earlier model LTM) Parts required for replacement: Tools required:   None Phillips screwdriver with torque meter  7/64” Allen wrench To temporarily remove the LTM/LTV mounting bracket: Lay the disconnected ventilator on a clean dry surface. Use a 7/64” Allen wrench to remove the three mounting block screws and mounting block, as shown below.
  • Page 145: Ltm/Ltv Mounting Bracket Permanent Removal (Earlier Model Ltm)

    LTM/LTV mounting bracket permanent removal (earlier model LTM) Parts required for replacement: Tools required:   Replacement Screws Kit, P/N 11149 Phillips screwdriver with torque meter ® To permanently remove the LTM/LTV mounting bracket: ® Remove the LTM/ LTV mounting block and bracket (see instructions on page 8-23). Insert five 1/4”...
  • Page 146: Ltm/Ltv ® Mounting Bracket, Reinstallation (For Earlier Model Ltm)

    LTM/LTV mounting bracket, reinstallation (for earlier model LTM) ® Parts required for replacement: Tools required:   Replacement Screws Kit, P/N 11149 Phillips screwdriver with Replace if damaged: torque meter  ®  LTM/ LTV mounting bracket, P/N 11125 7/64” Allen wrench ...
  • Page 147: Ltm/Ltv ® Mounting Bracket New Installation (Earlier Model Ltm)

    ® Contained in CareFusion LTM to LTV Mount Assembly Kit, P/N 11003. See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for additional information.
  • Page 148 WARNING Bracket screw location –Use of incorrect screws can cause ventilator failure with resulting harm to the patient. Insert five #4-40 3/8” back panel mounting screws (P/N 10879) and torque tighten to 60 in-oz (0.42 Nm). You may need to apply some pressure to the back panel and housing to correctly align the threads and insert the screws.
  • Page 149: Back Panel

    Back Panel Back Panel, Removal ® 1) If installed, remove the LTV protective boots (see instructions on page 8-5) and the ® LTM/ LTV mounting bracket (see instructions on page 8-23). 2) Turn the ventilator face down. Remove the 6 flat-head screws and gray finish washers from the right and left sides of the ventilator as shown (3 from each side).
  • Page 150: Back Panel, Reinstallation

    If the serial numbers do not match, locate the correct back panel for the ventilator before continuing. Contact CareFusion for back panel replacement information and proper labeling information. Sampling Tube, P/N 10544, ~10.0” long, 0.125” O.D. X 0.079 I.D. clear polycarbonate tubing.
  • Page 151 Verify the thermo conductive pad is applied to the inside of the back panel. If it is not, apply one. See instructions on page 8-130 to install. If necessary, remove the spiral wrap from the flow valve’s Stepping Motor 4-wire lead wires.
  • Page 152 NOTE An O Sampling Tube used for testing internal oxygen enrichment (oxygen leakage) during General Checkout testing, must be installed before replacing the ventilator’s back panel. See General Checkout in Chapter 9 – Final Checkout Test for test requirements. 16) Route an O Sampling Tube as shown in the illustration;...
  • Page 153 17) Inspect the conductive silicon gasket located on the outside surface of the top of the back panel where the three 1/4” screws secure the back panel to the ventilator. If the gasket is damaged or missing, replace it with a new one. 18) If the internal battery cable is disconnected, reconnect the battery cable to the power board.
  • Page 154: Internal Flexible Tubing

    CareFusion recommends that technicians review and become familiar with the existing tubing configuration prior to removing individual tubes and strictly adhere to the replacement instructions provided when reattaching them (point to point and layering).
  • Page 155: Tubing Table

    Tubing Table Tubing Material Diameter Length Installed Location /Color 11834 or One between elbow fitting from Silicone 0.188” (0.005m) O.D. 0.62” ( 0.1”) 18339-001 high pressure port on valve diff. /Clear 0.094” (0.002m) I.D. 0.016m transducer (PT7) & tee fitting and; One between port #7 on solenoid manifold &...
  • Page 156: Solenoid Manifold Tube Routing Table

    Tubing Material Diameter Length Installed Location /Color 10626 or Silicone 0.281” (0.007m) O.D. Between elbow fitting & flow diff. 0.85” ( 0.1”) 18338-001 transducer (PT4) /Clear 0.156” (0.004m) I.D. 0.022m 10626, Silicone 0.281” (0.007m) O.D. Between elbow fitting on turbine 10.9”...
  • Page 157: Solenoid Mount Tube Routing Table

    Solenoid Mount Tube Routing Table Tubing Solenoid Port Port Tubing Routing Pilot-in Through a filter subassembly onto the tee fitting in a tee & elbow subassembly attached to Solenoid Manifold port # 1 18336-005 Pilot-in Middle Through a wye subassembly onto the tee fitting in a tube & tee subassembly attached to Solenoid Manifold port # 9 18336-004 Pilot-in...
  • Page 158: Tubing Removal/Replacement Instructions

    Tubing Removal/Replacement Instructions Parts required for replacement: Tools required:   None Grounded anti-static wrist strap Replace if damaged:  Measuring tool (Calipers, Ruler, etc) 0” thru 11” range   Flexible tubing Pliers - long round nose, smooth jaws – ESD safe ...
  • Page 159  the Solenoid Mount (p/n 18607-001) and Pilot Solenoid/Power Board Cable subassemblies (p/n 18530- 001) with attached tubing only needs to be disassembled when it is necessary to replace individual tubes or components within the subassemblies themselves To replace internal flexible tubing: ®...
  • Page 160 Solenoid Manifold Preparation: Attach tubing to the Solenoid Manifold P/N 18528-001 ports;  Port # 2 – 4.70” of tubing P/N 10458  Port # 3 – 3.25” of tubing P/N 11834  Port # 4 – 1.75” of tubing P/N 11834 ...
  • Page 161 Construct a Tube and Tee subassembly, as follows;  Insert Tee fitting P/N 18186-001 into 2.06” and 1.75” of tubing P/N 11834  Attach the open end of the 1.75” tubing to manifold side port # 9 Attach the Tee & Elbow subassembly to solenoid manifold port # 1, oriented as shown, with;...
  • Page 162 Solenoid Mount Preparation: Construct a Plug subassembly, as follows:  fully insert Plug fitting P/N 18187-001 into 0.85” tubing P/N 10456 Construct a Wye subassembly, as follows:  attach a 1.00” tubing P/N 10456 to each top end (2) of Wye fitting P/N 18185-001 and a 0.50”...
  • Page 163 Attach subassemblies to the Solenoid Mount assembly, as follows:  attach the open end of the Plug subassembly tube to the bottom port of the Pilot-in solenoid on the Solenoid Mount P/N 18607-001  attach the end of the Wye subassembly tube with the installed restrictor onto the middle port of the Pilot-in solenoid (restrictor will slide farther into tube during attachment) ...
  • Page 164 Analog PCBA Preparation: As applicable, prepare the Analog PCBA P/N 18242-001 for installation, as follows:  attach 0.50” of tubing P/N 10456 to the high and low pressure ports (onto at least two-thirds of the height of the fittings) on the Valve Differential Transducer ...
  • Page 165 Solenoid Manifold, Analog PCBA and External Port Tubing Installation: NOTE Some previously attached manifold tubing is not shown for clarity of pertinent instructions. Attach Solenoid Manifold port tubing, as follows;  port # 4 onto ventilator side panel Low Pressure port ...
  • Page 166 Solenoid Mount and Subassemblies Installation: Position the Manifold to Back Panel Spacer P/N 18142-001 on the Turbine Manifold, align the mounting holes, insert four 7/16” pan-head screws and torque tighten each to 20 in-oz (0.14Nm). Position the Purge Solenoid/Power PCBA cable connectors temporarily out of the way (as shown) and route the 4-wire cable on the Pilot Solenoid/Power PCBA cable assembly under the Accumulator mounting tab on the Manifold to Back Panel Spacer and between the Purge Solenoids and the Turbine Manifold.
  • Page 167 Align and lower the Solenoid Mount assembly into the cavity between the Solenoid Manifold and the Flow Valve assembly.  as the assembly is lowered into position, loop and position the excess Pilot Solenoid / Power PCBA cable as shown in the illustration Attach the Solenoid Mount assembly to the end of the Manifold to Back Panel Spacer using the screw (P/N 17682-103, 4-24 X 3/8”...
  • Page 168 Connect the open end of the Wye subassembly tubing onto the tee fitting in the Tube and Tee subassembly attached to Solenoid Manifold port # 9. Connect the open end of the Filter subassembly onto the tee fitting in the Tee and Elbow subassembly attached to Solenoid Manifold port # 1.
  • Page 169 Accumulator Installation: Attach a 3.70” long precut silicone tube (P/N 10458) onto the PEEP Pilot Pressure Transducer port. Position the balance of the tube beside the Purge Solenoids, over previously installed tubing, and under the Accumulator mounting tab on the Manifold to Back Panel Spacer, as shown.
  • Page 170 Position and hold the Accumulator above the mounting tabs on the Manifold to Back Panel Spacer, as shown. Attach the open end of the Tube and Tee subassembly from Solenoid Manifold port # 9 onto port A of the Accumulator. Attach the open end of the 3.70”...
  • Page 171 Final Tubing and Cable Installation: Route the tube from Solenoid Manifold port # 3 under the tube and tee attached to Solenoid Manifold port # 7, over previously installed tubing and onto the Airway Pressure Transducer port, as shown. Route the tube from Solenoid Manifold port # 2 under the tube and tee attached to Solenoid Manifold port # 7, over the previously attached tubing, through the notch in the Analog PCBA and onto the Flow Differential Transducer port, as shown.
  • Page 172 Attach the Pisco connector onto the O Pressure Transducer port on the Analog ® PCBA, as shown (LTV 1200 only). Insert the tube from the O Blender straight into the top of the Pisco connector (at ® least 1/2”), as shown (LTV 1200 only).
  • Page 173 Route the 2-wire, Thermistor cable from the Flow Valve over previously installed tubing,  through the notch in the Analog PCBA,  between the Analog PCBA and the capacitors on the Power PCBA and  under the tube attached to the Flow Differential Transducer. Attach the connector to the 2-wire connector (JP8) on the Power PCBA, as shown.
  • Page 174: Accumulator

    Accumulator Parts required for replacement: Tools required:   Phillips screwdriver with torque meter Accumulator P/N 18144-001  Grounded anti-static wrist strap Replace if damaged:  7/16” Pan Head screws (2) P/N 10433 To remove the accumulator: Remove the ventilator back panel and disconnect the internal battery cable from the power board (see instructions on page 8-28).
  • Page 175 ® Page 8-54 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 176: Alarm Sounder Assembly (Previous Version)

    Alarm Sounder Assembly (previous version) Parts required for replacement: Tools required:   Sounder Bracket P/N 10119 Phillips screwdriver with torque meter  Sounder Assembly P/N 17432-001  Grounded anti-static wrist Replace if damaged: strap  Alarm Bumper P/N 10573 ...
  • Page 177 6) Place the sounder assembly in the cut-out in the right side of the upper weldment. 7) Line the sounder bracket up with the sounder and the mounting hole in the side panel. Orient the sounder assembly so that the slot in the grounding clip aligns with the hole in the sounder bracket, as shown.
  • Page 178: Alarm Sounder Assembly (Current Version)

    Alarm Sounder Assembly (current version) Parts required for replacement: Tools required:  Sounder Assembly P/N 21150-001  Phillips screwdriver with (contained in Sounder Kit P/N 21677-001) torque meter Replace if damaged:  Grounded anti-static wrist  1/4" Pan-head Screw (1) P/N 10435 strap To replace the Sounder Assembly: 1) Remove the ventilator back panel and disconnect the internal battery cable from the...
  • Page 179 5) Install the current sounder assembly in the cut-out in the right side of the upper weldment. 6) Line the sounder assembly up with the mounting hole in the side panel as shown. 7) Replace the pan-head screw in the sounder bracket and torque tighten to 60 in-oz (0.42 Nm).
  • Page 180: Analog Board Assembly

    Analog Board Assembly Parts required for replacement: Tools required:   Analog PCBA Assembly P/N 18242-001 Phillips screwdriver with torque meter  Replace if damaged: Grounded anti-static wrist strap  Pisco connector P/N 10543 To replace the analog board: 1) Remove the back panel of the ventilator and disconnect the internal battery cable (see instructions on page 8-28).
  • Page 181 3) Disconnect the 4-wire flow valve connector from the motor board. 4) Disconnect the Pisco connector from the oxygen pressure transducer. To do this, pinch the collar on the Pisco connector away from the Analog board as shown ® (LTV 1200 only).
  • Page 182 Analog board installation: 1) Insert the new analog board and secure with the two green mounting screws. Screws should be torqued to 60 in-oz (0.42 Nm). ® 2) Reconnect the Pisco connector to the oxygen pressure transducer (LTV 1200 only). 3) Reattach the 6 flexible tubes to the new analog PCBA following the internal flexible tube routing tables, diagrams and instructions on page 8-33.
  • Page 183 4) Reconnect the 4-wire flow valve connector to the motor board. The connector is keyed to fit in only one way and will snap into place when properly aligned. 5) Reconnect the 2-wire flow valve thermistor cable to the power board. 6) Inspect all flexible tubes for tears at the connecting ends and replace worn or damaged tubes if necessary.
  • Page 184: External Inlet Filter

    External Inlet Filter The external inlet filter should be removed and cleaned once a month. If the ventilator is being operated in high dust or humidity environments, it may need to be cleaned more often. If the filter is damaged or can not be thoroughly cleaned, it should be replaced. Parts required for replacement: Tools required: Replace if damaged:...
  • Page 185: Fan Assembly

    Fan Assembly Parts required for replacement: Tools required:   Phillips screwdriver with torque meter Fan Assembly P/N 10675, or 14725-  Grounded anti-static wrist strap Replace if damaged:  5/8" Flat-head Screw (2) P/N 10499  Nut (2) P/N 10342 To replace the fan assembly: 1) Remove the back panel of the ventilator and disconnect the internal battery cable (see page 8-28).
  • Page 186: Cleaning The Fan Filter

    Cleaning the Fan Filter The fan filter should be removed and cleaned at least once a month. If the ventilator is being operated in high dust or humidity environments, it may need to be cleaned more often. If the filter is damaged or can not be thoroughly cleaned, it should be replaced. To clean the fan filter Using a small screwdriver or long nose pliers, detach the fan filter grill from its housing.
  • Page 187: Flow Valve Assembly And Cleaning Instructions

    Flow Valve Assembly and Cleaning Instructions Parts required for replacement: Tools required:   Flow Valve Assembly P/N 10019 Phillips screwdriver with torque meter  Silicone Gel Lubricant P/N 10123  Grounded anti-static wrist strap  Thermistor Cable P/N 11399 ...
  • Page 188 6) Remove the 2 screws from the body of the flow valve. 7) Lift the flow valve up and out of the ventilator case. 8) Disconnect the bypass tubing from the connector at the base of the flow valve. Flow Valve Cleaning Procedure In certain cases, dust and other particulates may gather on the Poppet of the Flow Valve.
  • Page 189 FIGURE 1 4) Wrap the alcohol wipe around the Flow Valve Poppet as shown below in FIGURE 2, and pull forward and back several times to wipe the surface of the Flow Valve Poppet (See FIGURE 3 for detail of Poppet surface.) FIGURE 2 ®...
  • Page 190 FIGURE 3 5) To remove the alcohol wipe, lightly push the opposite (Step motor) end of the push rod while gently pulling on the wipe. This aids in removing foreign matter from the Flow Valve seat. 6) Using a fresh alcohol wipe, repeat Steps 1 through 5 several times, or more if needed, to remove all substances from the Flow Valve Poppet surface.
  • Page 191 Flow Valve preparation: 1) If replacing the same Flow Valve, skip to the next section. 2) If installing a new Flow Valve, P/N 10019, measure the length of the clear silicone tube attached to the port above the orange silicone tube (see illustration). ...
  • Page 192 To replace the flow valve assembly: Connect the bypass tubing to the connector at the base of the new flow valve. Apply a small amount of silicone gel lubricant to the flat surface area on the valve sealing gasket. Apply a small amount of silicone gel lubricant to the Mylar sheet insertion tool.
  • Page 193 Carefully route the Thermistor Cable between the tubing, through the notch in the analog PCBA, and connect it to JP8 on the power board as shown. The cable should tuck behind the capacitors on the power board to avoid strain. NOTE Make certain the Thermistor Cable does not interfere with the flow valve operation and that it does not strain the tubing connected to the solenoid manifold and analog PCBA.
  • Page 194 Loop the bypass tubing down into the space between the manifold and the bottom edge of the upper weldment as shown so that it will not be pinched when the back of the ventilator is replaced. Connect the orange tube to the valve differential transducer low pressure port on the analog board (furthest from the board).
  • Page 195 15) Select the VHome setting printed on the VHome label of the flow valve. 16) Enter the TEMP COMP menu and make sure that temperature compensation is set to TCOMP ON. 17) Power the ventilator off and set dip switch 5 to the OFF position (see instructions on page 6-5).
  • Page 196: Front Panel

    Front Panel Parts required for replacement: Tools required: ®   1200 Membrane Switch Panel Phillips screwdriver with torque meter P/N 17513-001  ®  Dental pick or flat tip screwdriver 1150 Membrane Switch Panel  P/N 17513-006 Grounded anti-static wrist strap ®...
  • Page 197 10) Remove the accumulator from the turbine manifold spacer. 11) Remove the four (4) screws securing the turbine manifold spacer to the turbine manifold assembly. Loosen or remove the single screw securing the solenoid mount assembly and remove the two screws holding down the accumulator. Move the solenoid mount assembly and the accumulator from the spacer.
  • Page 198 CAUTION Damage to contacts - Handle the ribbon cable carefully to prevent any damage to the silver contact area. Attempting to remove or attach the ribbon cable without opening the ZIF connector will damage the cable and may require replacement of the front membrane panel. 13) Slide the ZIF connector on the keypad ribbon cable open as shown below and carefully remove the ribbon cable from the connector.
  • Page 199 Installing a new front panel assembly: Remove the protective backing from the bottom side only of the new membrane panel. Hold the upper weldment upright against a well-lit backdrop to assist in aligning the membrane panel with the cutouts in the upper weldment. Handle the ribbon cable carefully to prevent any damage to the silver contact area.
  • Page 200 ® 10) Reconnect the 8-wire oxygen blender connector to the power board (LTV 1200 only) . 11) Ensure that the flexible tube from port B on the accumulator is tucked between the purge solenoids and the turbine manifold. Reattach the turbine manifold spacer to the turbine manifold using the four (4) 7/16”...
  • Page 201 20) Install an O Sampling Tube . See Back Panel, Reinstallation in Chapter 8 – Component Removal and Replacement for detailed instructions. 21) Verify that no tubing is kinked or twisted, and that it will not be pinched when compressed by the back panel.
  • Page 202: Internal Battery Pack

    Internal Battery Pack Parts required for replacement: Tools required:  LTV ®  1200, 1150, and 1100 Internal Torque driver (120 in-oz/0.84 Nm) Replacement Battery P/N 18634-001.  3/32” Hex bit CAUTION ® Internal Battery Use - The internal battery is intended for use in LTV ventilators during short periods while switching between external power supply connections, emergency situations or short duration transports.
  • Page 203 3) Lift the battery out of the battery compartment and while moving it over to sit on the open battery cover, gently pull on the battery wires to remove the battery connector and cable from their protective enclosures. 4) Press the connector release latch, disconnect the battery wire connector and set the original battery aside.
  • Page 204 7) Ensure that the battery connector and battery cable are fully inserted into their enclosures, tilt the battery as shown and begin inserting it into the battery compartment. 8) Straighten the battery and continue pressing down until it is fully seated in the battery compartment.
  • Page 205: Caring For The Internal Battery

    Charge Status LED is not illuminated green within 24 hours, or if it is illuminated red, immediately contact a certified service CareFusion technician or CareFusion.  Store the ventilator at temperatures less than 60°C (140°F).  A new battery should be charged on standby for 24 hours prior to use on a patient.
  • Page 206: Internal Inlet Filter

    Internal Inlet Filter The interior inlet filter should be removed and cleaned every 2 years or 10,000 hours. If the filter is damaged or can not be thoroughly cleaned, it should be replaced. Parts required for replacement: Tools required:  Phillips screwdriver with torque meter Replace if damaged: ...
  • Page 207: Main Board Assembly

    Main Board Assembly Parts required for replacement: Tools required:   Main PCBA Assembly P/N 18233-001 Phillips screwdriver and hex driver or P/N 14791-001 with torque meter  Replace if damaged: Grounded anti-static wrist strap  1/4" Pan-head Screw (3) P/N 10435 ...
  • Page 208 Remove the two (2) screws that secure the accumulator to the turbine manifold spacer, and remove the accumulator from the spacer. 10) Remove the four (4) screws that secure the turbine manifold spacer to the turbine manifold assembly. 11) Remove or loosen the one (1) screw that secures the solenoid mount assembly to the turbine manifold spacer, remove the solenoid mount assembly from the spacer, and set spacer aside, away from the ventilator.
  • Page 209 14) Remove the 3 main board mounting screws and the two hex standoffs. 15) Hold the ribbon cable out of the way and remove the main board. CAUTION Damage to the ribbon cable - Do not scratch the ribbon cable with the edge of the board even minor scratches on the connections can cause the keypad connections to fail.
  • Page 210 17) Hold the ribbon cable out of the way and install the new main board. Be careful not to scratch the ribbon cable with the edge of the board. When the main board is correctly aligned, the LEDs and displays will pop into place in the cutouts in the upper weldment.
  • Page 211 26) When LEDs are aligned, return all DIP switches to the default positions. 27) Reconnect the flexible tubes to the Flow Xducer and Exh Valve fittings on the side panel of the ventilator following the diagrams, tables and instructions beginning on page 8-33.
  • Page 212 30) Ensure that the flexible tube from port B of the accumulator is tucked between the purge solenoids and the turbine manifold. Replace the turbine manifold spacer using four (4) 7/16” pan-head screws (P/N 10433), Torque-tightened to 60 in-oz (0.42 Nm). 31) Replace the accumulator (see Accumulator on page 8-53).
  • Page 213: Memory Board

    Memory Board The memory board contains the software that operates the ventilator. The memory board should need to be replaced only when upgrading the software to a new version. It is a generally a good practice to upgrade the ventilator software to the most current released version at the time the ventilator is being serviced.
  • Page 214 Installing the new memory board: 1) Verify jumper is installed in the Normal Operation position of the JP3 connector (as shown), prior to installing the memory board. 2) Install the new memory board by lining the edges up in the supporting rails and sliding it into position. Be sure the connector on the memory board is oriented correctly to line up with the mating connector on the main PCBA.
  • Page 215: Motor Board Assembly

    Motor Board Assembly Parts required for replacement: Tools required:   Motor PCBA Assembly P/N 10135-002 Phillips screwdriver with torque Replace if damaged: meter   7/16" Pan-head Screw (4) P/N 10433 Grounded anti-static wrist strap To replace the Motor Board 1) Remove the back panel of the ventilator and disconnect the internal battery cable (see instructions on page 8-28).
  • Page 216 Installing the motor board: 1) Verify the power board mating connector pins are straight and undamaged. 2) Install the new motor board by placing it on the four threaded stand-offs on the power board. Align the board so that the connector pins on the power board are indexed into the mating connector holes in the motor board.
  • Page 217: Blender Assembly / O Block

    Blender Assembly / O Block Parts required for replacement: Tools required:   Blender Assembly P/N 15079-001 or O Phillips screwdriver with torque meter Block with Fittings P/N 18345-001  Grounded anti-static wrist strap  Cable Tie P/N 10466  Small dykes or cutters ...
  • Page 218 To remove the Turbine Manifold and Oxygen blender/O Block: Remove the back panel and disconnect the battery (see instructions on page 8-28). Remove the left soft side (see instructions on page 8-117). ® Remove the motor board (LTV 1200 only) (see instructions on page 8-94). ®...
  • Page 219 Remove the four (4) black screws from the turbine manifold. Remove the four (4) blender mounting screws and metal washers from the left side of the ventilator. 10) Remove the oxygen blender/O Block, turbine manifold and interior inlet filter. Be careful not to dislodge the grounding straps and grommets that are between the oxygen blender and the side of the ventilator case.
  • Page 220 To connect the Turbine Manifold to a new Oxygen Blender/O Block: Handle the manifold and blender/O Block carefully so as not to break the silicone seal around the oxygen tube entering the manifold. If this seal is damaged, repair with RTV silicone adhesive.
  • Page 221 Check the blue rubber grommets (P/N 10630) and grounding clips (P/N 10752) on the ® oxygen blender mounting (4 each) for wear and replace if necessary (LTV 1200 only). ® For the LTV 1200: Replace the four (4) blender mounting screws (P/N 10176), O blender cap (P/N 15016-001 or 13301-001), and four (4) flat metal washers (P/N 10594).
  • Page 222 Connect the 8-wire connector from the oxygen blender to the power board. The connector is keyed to fit in only one direction and will snap into place when properly ® connected (LTV 1200 only). Tuck the wrapped wires and oxygen tube down along the side of the turbine manifold ®...
  • Page 223 12) Connect the bypass tubing from the flow valve to the barbed elbow at the bottom of the turbine manifold. The tubing should be looped into the space between the bottom of the manifold and the bottom edge of the upper weldment so that it is out of the way and will not be pinched when the back of the ventilator is replaced.
  • Page 224 16) Reconnect the orange tube from the flow valve to the valve differential transducer low pressure port on the analog board (furthest from the board). 17) Reconnect the clear flexible tube from the flow valve to port #5 on the solenoid manifold.
  • Page 225 Blender Filter (LTV 1200 only) ® The O inlet filter should be cleaned or replaced every 10,000 hours or 2 years of service or more frequently if needed. If the ventilator is operated with a low grade or contaminated O source, the O inlet filter may need to be replaced more often.
  • Page 226: Power Board Assembly

    Power Board Assembly Parts required for replacement: Tools required:   Power PCBA P/N 18867-001, Phillips screwdriver with torque meter P/N 18120-001, P/N 14944-001,  Grounded anti-static wrist strap or P/N 21330-001  P/N 11599 – Power PCB Separator  1/4”...
  • Page 227 Remove or loosen the one (1) screw with washer that secures the solenoid mount assembly to the spacer. Remove the two (2) screws attaching the accumulator to the spacer and remove the accumulator from the turbine manifold spacer. 10) Remove the four (4) screws that secure the turbine manifold spacer to the turbine manifold.
  • Page 228 13) Remove the power board mounting screws (2 attaching the power board and/or the Pigtail cable assembly to the side panel from the outside (retain) and five (5) located on the face of the board). See illustration below. 14) Pull the pigtail connector interface bracket straight up and off the power board connectors and the molded strain relief portion of the pigtail cable assembly.
  • Page 229 18) Position the back surface of the lower portion of the tool against the turbine manifold with the lowest edge of the tool resting on the main PCBA below the power board. Hold the lower portion of the tool against the turbine manifold and rotate the handle backwards and down toward the top surface of the turbine manifold to separate the power and main PCBA mating connectors (see Cut-away illustration below).
  • Page 230 NOTE The solenoid mount assembly, accumulator, and remaining tubing may be left attached to the solenoid manifold assembly. 20) Remove the solenoid manifold from the power board. ® p/n 18603-001, Rev. J 1200, 1150, and 1100 Ventilator Service Manual Page 8-109...
  • Page 231 Installing a new power board: Place the power board over the main PCBA. Align the two external connectors (patient assist and comm. port) with the two square holes in the side of the upper weldment and align the 64-pin female connector on the power PCBA with the 64-pin male connector on the main PCBA (see illustration).
  • Page 232 Slide the solenoid manifold. over the threaded mounting stud on the power board. Align the solenoid leads with the JP17 connector on the power board being installed and press the solenoid manifold into place. Insert a 7/8” red colored pan-head solenoid manifold mounting screw (P/N 10607R), as shown and torque-tighten to 20 in-oz (0.14 Nm).
  • Page 233 10) Insert the cylindrical connector-end of a pigtail cable assembly over the power board and through the keyed-round hole of the upper weldment from the inside of the upper weldment. 11) Slide the pigtail cable assembly through the hole in the upper weldment until the strain relief protrudes through the keyed-round hole in the upper weldment.
  • Page 234 15) Reconnect the 8-wire blender connector. Tuck the wrapped wires and oxygen tube ® down along the side of the turbine manifold against the power board (LTV 1200 only). 16) Verify the power board mating connector pins are straight and place the Motor PCBA on the four threaded stand-offs on the power board and orientated such that the connector pins on the power board are indexed into the matching connector holes in the Motor PCBA.
  • Page 235 NOTE When the Motor PCBA is correctly seat on the power board, the pins on the power board should visibly protrude through the top of the connector on the Motor PCBA. 18) Thread four 7/16” pan-head screws (P/N 10433) into the mounting holes in the Motor PCBA and torque-tighten to 60 in-oz (0.42 Nm).
  • Page 236 19) Replace the alarm sounder, (see instructions on page 8-55). 20) Reconnect the 2-wire fan connector, solenoid mount assembly 4-wire connector, 4-wire rotary switch connector and 2-wire sounder connectors to the power board. The connectors are keyed to fit in only one direction and will snap into place when properly connected.
  • Page 237 26) Reconnect the three (3) flexible tubes to the Flow Xducer and Exh Valve fittings on the side panel of the ventilator. 27) Reinstall the flow valve assembly (see instructions on page 8-66). 28) Reconnect the flexible tubes to the solenoid manifold following the internal flexible tube routing table, instructions and diagrams beginning on page 8-33.
  • Page 238: Right And Left Soft Side Panels

    Right and Left Soft Side Panels Parts required for replacement: Tools required:  Replace if damaged: Phillips screwdriver with torque meter  Left Soft Side (dark gray) P/N 10105  Grounded anti-static wrist  Right Soft Side (dark gray) P/N 10106 strap ...
  • Page 239: Rotary Switch (Set Value Knob) Assembly

    Rotary Switch (Set Value Knob) Assembly Parts required for replacement: Tools required:  Phillips screwdriver with torque  Rotary Switch Replacement Kit P/N 14271 meter  Rotary Switch (with hex nut) P/N 11190  2 small flat tip screwdrivers or ...
  • Page 240 11) Carefully remove and discard the rotary switch assembly washer and O-ring from the back of the ventilator through the openings in the power and main boards. 12) Place the O-Ring (P/N 11645) inside the Washer (P/N 11644), insert them through the openings in the power and main boards and center both over the Rotary Knob shaft cutout in the upper weldment.
  • Page 241: Solenoid Manifold Assembly

    Solenoid Manifold Assembly Parts required for replacement: Tools required:  Solenoid Manifold Assembly P/N 18528-001  Phillips screwdriver with torque meter  1/4” Nut Driver adapter for torque wrench  Grounded anti-static wrist strap  Torque screwdriver (120 in-oz/0.84 Nm) Solenoid Manifold Tube Routing Table Solenoid Tubing...
  • Page 242 To replace the Solenoid Manifold: Remove the back panel of the ventilator and disconnect the internal battery cable (see instructions on page 8-28). Disconnect the 4-wire flow valve connector from the motor board. Disconnect the Pisco connector ® from the analog board (LTV 1200 only) .
  • Page 243 16) Remove the solenoid manifold from the power board, disconnect the flexible tube from port #9, and separate the solenoid mount assembly from the solenoid manifold. NOTE Tubing Configurations - When disconnecting tubes, note their positions and review the internal flexible tube routing tables, instructions and diagrams beginning on page 8-33. If tubing needs to be replaced, consult the tables and instructions in Internal Flexible Tubing beginning on page 8-33 for tubing diameters, lengths and routing diagrams.
  • Page 244 3) Slide the solenoid manifold over the threaded mounting stud on the power board, carefully align the solenoid contact pins with the JP17 connector on the power board and press the solenoid manifold into place. Make sure that the contact pins are correctly located in the JP17 receptacle.
  • Page 245 11) Referring to the diagrams and tables beginning on page 8-33, reattach the flexible tube from the tee and elbow subassembly, from port #1 on the solenoid manifold to the high pressure port of the flow valve differential transducer on the analog board (the port closest to the analog board).
  • Page 246 22) Reconnect the 4-wire flow valve connector to the motor board and the 2-wire flow valve Thermistor cable to the power board. The connectors are keyed to fit in only one direction and will snap into place when properly connected. 23) Install an O Sampling Tube .
  • Page 247: Solenoid Mount Assembly

    Solenoid Mount Assembly Parts required for replacement: Tools required:   Solenoid Mount Assy. P/N 18607-001 Phillips screwdriver with torque meter Replace if damaged:  Grounded anti-static  (4-24 X 3/8 Pan Head ) Screw P/N 17682-103 wrist strap  (.125 I.D.
  • Page 248 6) Loosen or remove the one (1) screw that secures the solenoid mount assembly to the turbine manifold spacer, and remove the solenoid mount assembly from the spacer. NOTE When removing the tubes from the pilot-in solenoid, make sure that the small brass filter and restrictor remain in the tubes.
  • Page 249 3) Refer to page 8-45 for detailed instructions, and attach the solenoid mount assembly to the turbine manifold spacer. 4) Reconnect the two 2-wire connectors to the purge solenoids. Orientation does not matter. 5) Reinstall the flow valve assembly (see instructions on page 8-66). 6) Reconnect the 3-wire flow valve connector, to the power board.
  • Page 250: Thermo Conductive Motor Board Heatsink Pad

    Thermo Conductive Motor Board Heatsink Pad Thermo conductive pads should be replaced any time they have hardened, and at 30,000 hours. The motor board heat sink pad is located on the inside of the back panel. Parts required for replacement: Tools required: ...
  • Page 251: Thermo Conductive Turbine Pad

    Thermo Conductive Turbine Pad Thermo conductive pads should be replaced any time they have hardened, and at least at the 30,000 hour service. The turbine heat sink pad is located between the inside of the upper weldment and the turbine. Parts required for replacement: Tools required: ...
  • Page 252: Turbine Manifold

    Turbine Manifold Parts required for replacement: Tools required:   Turbine Manifold Assembly P/N 11490 Phillips screwdriver with torque  meter Cable Tie P/N 10466   Grounded anti-static wrist strap Donut Seal P/N 10603   Small dykes or cutters Silicone Gel Lubricant P/N 10123 ...
  • Page 253 To remove the Turbine Manifold and Oxygen blender or O Block: Remove the back panel and disconnect the battery (see instructions on page 8-28). Remove the left soft side (see instructions on page 8-117). Remove the motor board (see instructions on page 8-94). ®...
  • Page 254 Remove the four (4) black screws from the turbine manifold. Remove the four (4) blender / O Block mounting screws and washers from the left side of the ventilator. ® ® 10) Remove the oxygen blender (LTV 1200 ) or O Block (LTV 1150, and 1100), turbine manifold and interior inlet filter.
  • Page 255 To connect the new Turbine Manifold to the Oxygen Blender or O Block: Handle the manifold and blender carefully so as not to break the silicone seal around the oxygen tube entering the manifold. If this seal is damaged, repair with RTV silicone adhesive.
  • Page 256 Replace the 4 black pan-head screws (P/N 10918B) into the turbine manifold. Screws should be torqued to 20 in-oz (0.14 Nm). Check the blue rubber grommets (P/N 10630) and grounding clips (P/N 10752) on the ® oxygen blender mounting (4 each) for wear and replace if necessary (LTV 1200 only).
  • Page 257 NOTE If the upper clips are not properly aligned with the edge of the side panel, the back panel of the ventilator will not seat correctly. Use a straight edged screwdriver or pick to prevent each ® grounding clip from turning as the mounting screws are tightened (LTV 1200 only).
  • Page 258 12) Connect the bypass tubing from the flow valve to the barbed elbow at the bottom of the turbine manifold. The tubing should be looped into the space between the bottom of the manifold and the bottom edge of the upper weldment so that it is out of the way and will not be pinched when the back of the ventilator is replaced.
  • Page 259 16) Reconnect the orange tube from the flow valve to the valve differential transducer low pressure port on the analog board (furthest from the board). 17) Reconnect the clear flexible tube from the flow valve to port #5 on the solenoid manifold.
  • Page 260: Chapter 9 - Final Checkout Test

    The LTV 1100 Ventilator shall be tested while in the Flow Sensing Patient Circuit Mode (FL SENSOR). The patient circuit mode Default Setting is Flow Sensor. Contact CareFusion or ® refer to Chapter 12 – Operating Procedure in the LTV 1100 Operators Manual P/N 21106-001 for detailed information.
  • Page 261: General Checkout

    General Checkout Tools required:  AC Adapter  Calibrated O Analyzer  Compressed O source with a 0-80 PSI (0-5.51 bar) regulator  Patient Circuit, Adult, without PEEP  1 liter or greater test lung with compliance of 10ml/cmH O, resistance of 5 cm/L/sec ...
  • Page 262 NOTE When the Extended Features PRESETS, PTNT QUERY menu option is set to QUERY ON and the ventilator is powered up in normal ventilation mode, ventilation and alarm activation are suspended and the message SAME PATIENT is displayed after completing POST tests. Press the Select button to enable the suspended alarms and begin ventilation with the settings in use during the last power cycle, or set the PTNT QUERY menu option to QUERY OFF.
  • Page 263 B) Set the O % control to 22% LOW O2 PRES alarm activates ® (LTV 1200 only). C) Reset O % to 21% and clear the LOW MIN VOL alarm activates ® alarm (LTV 1200 only). Set the Low Min Vol alarm to 10 L. D) Reset the Low Min Vol alarm to LOW PRES alarm activates 1.0 and clear the alarm.
  • Page 264 L) Return to normal ventilation POWER LOST alarm activates. mode . Disconnect AC Adapter Battery Level LED illuminates showing the from ventilator. charge level. Ventilator continues to operate from the internal battery. M) Clear the POWER LOST alarm. After the ventilator is turned off, the Vent Turn the ventilator off.
  • Page 265 H) Turn the ventilator back on in Resistance > 1.0 mega ohms breathing mode, clear all alarms and measure the plug resistance. I) Create a High Pressure alarm by Resistance ≤ 4.0 ohm changing the High Pres. Limit setting to 5 and measure the plug resistance.
  • Page 266 6) Check usage hours, date and time. A) Turn the ventilator on in breathing mode and access the extended features by pressing and holding the Select button until ALARM OP appears in the display window. B) Navigate to VENT OP, press the The displayed usage hours shall be Select button and navigate to greater than or equal to the received usage...
  • Page 267 NOTE may exhaust from Air Inlet filter Port and falsely elevate reading. Do not permit exhausted O to enter via the fan intake. 7) Perform an Internal Oxygen Enrichment (oxygen leakage) test. A) Connect an oxygen sensor with a Peak % O measurement taken from inside the unit must be ...
  • Page 268: General Checkout Worksheet

    General Checkout Worksheet ® ____________ ____________ SERIAL NUMBER: CONDUCTED BY: DATE: ____________ PASS TEST STEP MEAS. (P) / REQUIREMENT FAIL DESCRIPTION VALUE Ventilator Checkout ("VENT CHECK") Alarm 2)A) Audible alarm tone must activate Confirming audible chirp must activate after alarm is silenced. Display 2)B) All displays must light except Vent Inop...
  • Page 269 Disc / Sense Alarm High side sense line disconnected 4)I) Alarm must activate at start of next from unit inspiration Pressure Operation (LTV 1200, 1150 only) Control Settings: 4)K) Monitors should read as follows: Pressure Assist / Control PIP: 46 to 54 cmH Breath Rate = 10 No alarms activate.
  • Page 270 ® SERIAL NUMBER: _____________ PASS TEST STEP MEAS. (P) / REQUIREMENT FAIL DESCRIPTION VALUE Patient Assist Port Connect Patient Assist Cable, Normally 5)A) Closed (P/N 10779), reconnect AC adapter, power ventilator up and clear all alarms. Measure plug resistance. 5)B) Resistance ≤...
  • Page 271 ® SERIAL NUMBER: _____________ PASS TEST STEP MEAS. (P) / REQUIREMENT FAIL DESCRIPTION VALUE Usage Hours, Date and Time Verification Power up ventilator and access 6)A) extended features. Display usage hours. 6)B) The displayed usage hours shall be greater than or equal to the received usage hours.
  • Page 272: Power Checkout

    Power Checkout Tools required:  AC Adapter  Internal Battery Test Cable , P/N 11472  External Battery Test Cable , P/N 11474  Variable 10 Amp DC voltage source capable of displaying the source output value to one hundredth of a volt ...
  • Page 273 2) Perform a Battery Duration/Battery Charge test  A) Operate unit from its internal battery BAT LOW, BAT EMPTY, and Vent Inop until it shuts down. Confirm Vent alarms sound Inop LED is lit and audible alarm is  When the unit shuts down confirm Vent Inop sounding.
  • Page 274 E) Remove external battery test cable and replace with external power AC adapter. Power down the ventilator using the On/Standby button. F) If further testing is not required with the back panel off, reinstall the back Back Panel, panel (see Reinstallation in Chapter 8).
  • Page 275: Power Checkout Worksheet

    Power Checkout Worksheet ® ____________ ____________ SERIAL NUMBER: CONDUCTED BY: DATE: ____________ PASS TEST STEP (P) / REQUIREMENT FAIL DESCRIPTION Battery Level Test Replace battery with the 10A DC 1)A) Normal operation voltage source set between 14.8V and Battery Level LED is green 15.0V and turn unit on.
  • Page 276 ® ____________ ____________ SERIAL NUMBER: CONDUCTED BY: DATE: ____________ PASS TEST STEP (P) / REQUIREMENT FAIL DESCRIPTION External Power Test With external AC adapter connected 3)A) Normal operation External Power LED is green Charge Status LED is lit or flashing Remove external power, clear power 3)B) Normal operation...
  • Page 277: Performance Checkout

    Performance Checkout Tools required:  AC Adapter  Gas Flow Analyzer  Turbine Pressure Test Adapter , P/N 11567  Oxygen Analyzer ®  0-80 PSIG O Supply (LTV 1200 only)  2 L test lung (or two 1 L test lungs in parallel) ...
  • Page 278 B) Measure breath rate or the interval must be 6.0 ± 0.5 sec. period between the start of two consecutive breaths. Perform a Pressure Control test (LTV 1200/1150 only). Set the ventilator as follows Pressure Control: 45 cmH O and...
  • Page 279  Status monitor displays “-“ (sensitivity Press the Sensitivity button and deactivated). rotate Set Value knob  Status monitor displays ALARM OP Press and hold the Select button  Status monitor window displays VENT OP Rotate the Set Value knob ...
  • Page 280 6) Perform a Real-time Transducer test. Turn the ventilator off. Turn the ventilator on while holding the Select, Control Lock, and Manual Breath buttons (3-finger mode ). Clear the REMOVE PTNT alarm, and then turn to the RT XDCR DATA menu and press Select.
  • Page 281 Turbine Pressure Test adapter from the ventilator. F) If the outlet pressure value is less than this value, troubleshoot (see Chapter 7 - Troubleshooting). For assistance, call CareFusion Technical Support using the contact information at the front of this manual.
  • Page 282 8) Perform a Servo test. Connect the ventilator to a high pressure oxygen source. Connect an oxygen sensor to the patient outlet port of the ventilator (for fastest response), or to the Patient Circuit Wye. NOTE ® Oxygen source and tested O % only apply to the LTV 1200 model.
  • Page 283 D) Turn the PPin solenoid off. Set  Leakage rate must be ≤ 1 cmH O within the turbine speed to 10 and one minute monitor the PPP transducer  The PPP value under the RT XDCR DATA reading. menu must be  0.5 cmH O of the measured pressure E) Turn the PPout solenoid off.
  • Page 284 11) Perform an LTM Compatibility test. If an LTM Monitor is not available to perform this test, downloading the Event Trace onto a computer using the Service Cable Assembly and Instructions contained in P/N 11485 (available seperatly or as part of the Maintenance and Calibration Kit, P/N 11566) will also test for the proper function of the communication port.
  • Page 285 12) Perform a Flow Valve Leak for I/E Hold. A) Power up ventilator in VENT MTNCE mode (i.e.: using the “three finger mode ” hold down the Control Lock, Manual Breath and Select buttons while powering up the ventilator) and enter VENT MNTCE.
  • Page 286: Performance Checkout Worksheet

    Period between start of 2 consecutive 2)B) Breath rate must be 10.0 ± 1.0 or breaths measured interval must be 6.0 ± 0.5 sec. Pressure Control (LTV 1200/1150 only) Settings: Pressure, Assist / Control, 3)A) PC= 45, PEEP= 5 Rise time= “as required”...
  • Page 287 ® SERIAL NUMBER: _____________ PASS (P) / TEST STEP MEAS. FAIL DESCRIPTION VALUE REQUIREMENT Sensitivity Settings: Volume, PEEP= 5, BPM=6, 4)A) TV=700, Insp Time=2.0, Sensitivity= 2 LEAK COMPENSATION OFF: 4)B) Observe the Patient Effort LED on the 4)C) The Patient Effort LED should upper right hand side of the ventilator momentarily flash and an inspiration control panel while creating a 4 Lpm...
  • Page 288 ® SERIAL NUMBER: _____________ PASS TEST STEP MEAS. (P) / REQUIREMENT FAIL DESCRIPTION VALUE Real-time Transducer Data Airway Pressure (AP): 0 cmH O to high flow XDCR port 6)A) AP display must read 0 ± 0.5 cmH 50 cmH O to high flow XDCR port 6)B) AP display must read 50 ±...
  • Page 289 ® SERIAL NUMBER: _____________ PASS TEST STEP MEAS. (P) / REQUIREMENT FAIL DESCRIPTION VALUE Solenoids Step position 100, turbine speed 4000 9)A) ExhPilot solenoid deactivated Pressure at exhalation drive port must = (OFF): 0 ± 0.1 cmH ExhPilot solenoid activated (ON): Pressure at exhalation drive port must be ≥14 cmH Step position 100, turbine speed 4000...
  • Page 290 ® SERIAL NUMBER: _____________ PASS (P) / TEST STEP MEAS. FAIL DESCRIPTION VALUE REQUIREMENT Watchdog Test Watch Dog activated CHECK: Unit shuts downs then restarts LTM Compatibility Test Set COM setting to MONITOR 11)A) Display of small LTV icon appears on the Connect the LTM serial com cable to the 11)B) ®...
  • Page 291: 12 Hour Burn-In

    12 Hour Burn-in Tools required:  AC Adapter  1 liter or greater test lung with compliance of 10 ml/cmH O and a resistance of 5 cm/L/sec OR a test lung setup with:  Mask Elbow 22mm/15F X 15M (Qosina P/N 51046) ...
  • Page 292 Front Panel Control Settings (using Test Lung P/N 18159-001, Air Restrictor P/N 19080-001, and Mask Elbow Qosina P/N 51046) LTV 1200, and 1150 LTV 1100 Control Setting Control Setting Breath Rate 15 bpm Breath Rate 15 bpm Tidal Volume 800 ml...
  • Page 293: 12 Hour Burn-In Worksheet

    12 Hour Burn-in Worksheet ® ____________ ____________ SERIAL NUMBER: CONDUCTED BY: DATE: ____________ PASS TEST STEP (P) / REQUIREMENT FAIL DESCRIPTION 12 Hour Burn-in Connect AC adapter and test lung. Power ventilator up and set the ventilator controls to the settings listed. Run the ventilator for a minimum of 12 Verify that no alarms occur.
  • Page 294 Appendix A - V ENTILATOR PECIFICATIONS Modes and Breath Types Breath Types Volume Control, Pressure Control, Pressure Support, Spontaneous Modes Control, Assist/Control, SIMV, CPAP, NPPV, Apnea Backup Variable Controls Control Range Tolerance Backup Pressure Trigger -3 cmH  2 cmH ...
  • Page 295 Control Range Tolerance Patient Query On, Off PEEP/CPAP 0 to 20 cmH Uncalibrated PIP LED Display On, Off Pressure Control 1 to 99 cmH  2 cmH O or 8% whichever is greater, steady-state only Pressure Control Flow On, Off Termination Pressure Support “--”, 1 to 60 cmH...
  • Page 296 Control Range Tolerance SBT High f 15-80 bpm  1 bpm or within 5%, whichever is greater  1 bpm or within 5%, whichever is SBT Low f 0-40 bpm greater Display f/Vt On - Off Alarms Variable Alarms Control Range Tolerance Apnea Interval...
  • Page 297 Fixed Alarms Control Range Tolerance / Indicators Default Settings EEPROM problem detected DISC/SENSE Positive (exhaled) airway flow during first 200 ms of (Low Pressure Sense inspiration and exhaled tidal Line Disconnect) volume (Vte) of previous breath is more than 4000 ml ...
  • Page 298 Volume  5 dBA Alarm Volume 60 to 80 dBA at one meter Vent Inop Ventilator Inoperative Immediately upon a ventilator inoperative condition, the audible indicator will begin sounding with a steady tone and the Vent Inop LED shall illuminate. Depressing the alarm Silence Reset button will silence the audible indicator.
  • Page 299 Button Controls Display Function Control Lock Locks front panel controls, can be set to Easy or Hard unlocking Manual Breath Generates a machine breath Standby/On Puts ventilator in On or Standby state Low Pressure O Source Selects Low Pressure O Source ®...
  • Page 300 Displays Display Range Tolerance Airway Pressure -10 to 108 cmH  3 cmH O or 5%, whichever is greater Display Window 12 characters Patient Effort Green LED Vent Inop Red LED External Power Amber / Green LED Charge Status Red / Amber / Green LED Battery Level Red / Amber / Green LED Usage Meter...
  • Page 301 Storage and Operating Conditions (cont.) Specification Tolerance Oxygen Inlet DISS or NIST Connector Inlet Pressure Range 40 to 80 PSIG  2 PSIG ® (LTV 1200 Only) Tapered Tubing Connector Inlet Pressure Range 0 to 10 PSIG  2 PSIG Shock and Vibration The ventilator is designed to withstand shock and vibration in accordance with relevant requirements set forth in the following standards:...
  • Page 302 Power Feature Range Tolerance / Indicators Input Voltage 11 to 15 VDC External Power AC Adapter Input: 100 to 250 VAC, 50 to 60  2.5% Output: 13 VDC Full Power Voltage  11.5 V  2% Green LED Voltage < 11.0V and  9.5V ...
  • Page 303 Power (cont.) Internal Battery 12V sealed lead acid battery. 4.5Ah. Feature Range Tolerance / Indicators Full Power Green LED Medium Power Amber LED Low Power Red LED Charge Time Battery shall be capable of being >90% When external power is charged within 8 hours, from fully present, and the vent is discharged state to state indicated by...
  • Page 304 EMC and RF Environments The following tables are provided in compliance with 60601-1-2 © IEC:2001(E), and describe the ® tested EMC limitations of the LTV ventilators. Table 201 Guidance and manufacturer’s declaration – electromagnetic emissions ® The LTV ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the entilator should assure that it is used in such an environment.
  • Page 305 Table 203 - 60601-1-2 © IEC:2001(E) Guidance and manufacturer’s declaration – electromagnetic immunity ® The LTV ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator should assure that it is used in such an environment. Immunity Test IEC 60601 test level Compliance Level...
  • Page 306 Table 205 - 60601-1-2 © IEC:2001(E) Recommended separation distances between portable and mobile RF communications equipment and the ® ventilator. ® The LTV ventilator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ventilator can help prevent electromagnetic interference by maintaining a minimum distance ®...
  • Page 307 ® Page A-14 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 308: Appendix B - Glossary

    Appendix B - G LOSSARY EFINITION Alternating Current. Airway Circuit The airway tubing that connects the ventilator and the patient. Airway Pressure The airway pressure measured at the exhalation valve. A bar graph type display composed of 60 LEDs. This display shows the Airway Pressure real-time airway circuit pressure from –10 cmH O to 108 cmH...
  • Page 309 EFINITION Centimeters of water pressure. A unit of measure for pressure. Control Mode A mode of ventilation where the patient receives a fixed number of machine breaths. In Control Mode, patient triggers are not allowed. CPAP Continuous Positive Airway Pressure. A positive pressure continually applied throughout the breath cycle.
  • Page 310 EFINITION Leak Compensation Leak Compensation improves triggering when a circuit leak is present. Light Emitting Diode. An indicator that is illuminated on the Front Panel. Liters Per Minute. A unit of measure for flow. Machine Breath A volume or pressure breath that is initiated by the operator or the ventilator, and is controlled and cycled by the ventilator.
  • Page 311 EFINITION PreSet A feature allowing ventilator parameters to be “preset” for an infant, pediatric, or an adult patient. Pressure Control Breath A machine or assist breath where the circuit pressure is elevated to a user-set pressure for a user-set period of time. Pressure Control Breaths have an optional flow termination criteria.
  • Page 312 EFINITION Volume Control Breath A machine or assist breath where a user-set volume is delivered over a user-set time. Flow is delivered in a decelerating waveform where the peak and final flows are calculated so that the final flow is 50% of the peak flow.
  • Page 313 ® Page B-6 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 314: Appendix C - Service Record Form

    Any time service is performed on the LTV 1200, 1150, or 1100 ventilator, a Service Record form should be filled out and returned to CareFusion. This enables complete tracking of replacement part lots and allows CareFusion to maintain comprehensive service history records.
  • Page 315 This page left deliberately blank ® Page C-2 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 316 ® p/n 18603-001, Rev.J 1200, 1150, and 1100 Ventilator Service Manual Page C-3...
  • Page 317 ® Page C-4 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 318 Appendix D - E VENT RACE The Event Trace is a list of events recorded by the ventilator . These events may be normal conditions, such as turning the ventilator on or off, or alarm conditions such as HW FAULT or HIGH PRES.
  • Page 319 xx:eventdate 5) Press Select and is displayed. is the chronological number of the event occurrence. eventdate is the date of the event; the date of the first occurrence is displayed in the initial occurrence recordings and the date of the latest occurrence in the secondary occurrence recordings of the same type of event.
  • Page 320 Event Codes This section includes a list of the event codes that can be recorded in the Event Trace. Event Codes by Code # Code Event Name Event Associated Alarm VENT 1 Power on None VENT 0 Power off None HOUR MTR Set hour meter None...
  • Page 321 Code Event Name Event Associated Alarm AD MTCH1 ADC mismatch occurred HW FAULT AD MTCH0 ADC mismatch cleared HW FAULT SYNCER1 Stepper motor lost sync occurred HW FAULT SYNCER0 Stepper motor lost sync exited HW FAULT HOME ER1 Stepper motor home failure occurred HW FAULT HOME ER0 Stepper motor home failure exited...
  • Page 322 Code Event Name Event Associated Alarm Problem detected with primary and/or HW FAULT RAC ERR1 redundant audible alarm circuitry Recovery from problem detected with primary HW FAULT RAC ERR0 and/or redundant audible alarm circuitry SNDRERR1 Alarm sounder error HW FAULT SNDRERR0 Recovery from alarm sounder error HW FAULT...
  • Page 323 Event Codes by Event Name Event Name Code Event Associated Alarm AD MMTCH ADC mismatch HW FAULT AD MTCH0 ADC mismatch cleared HW FAULT AD MTCH1 ADC mismatch occurred HW FAULT ADC1 VAL AD mismatch primary channel fault value HW FAULT APNEA 0 Apnea mode exited APNEA...
  • Page 324 Event Name Code Event Associated Alarm HIGH f0 High breath rate alarm recovered HIGH f HIGH f1 High breath rate alarm occurred HIGH f HOME ER0 Stepper motor home failure exited HW FAULT HOME ER1 Stepper motor home failure occurred HW FAULT HOUR MTR Set hour meter...
  • Page 325 Event Name Code Event Associated Alarm STACK Stack overflow detected RESET SYNCER0 Stepper motor lost sync exited HW FAULT SYNCER1 Stepper motor lost sync occurred HW FAULT TBN ESTP Turbine emergency stop occurred HIGH PRES TBN HSTP Turbine Hold Stop occurred HIGH PRES TBN ISTP Turbine immediate stop occurred...
  • Page 326 Event Trace Data Definitions XDC FLT1 Four binary digits, ABCD, where A represents the Flow Differential narrow (FDn) transducer channel B represents the Flow Differential wide (FDw) transducer channel C represents the Flow Differential bi-directional (FDb) transducer channel D represents the Airway Pressure (AP) transducer 1 = fault, 0 = okay For example, 0100 represents a failed auto zero on the FDw channel.
  • Page 327 ® Page D-10 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 328: Conversion Factors

    Appendix E - R EFERENCE NFORMATION Conversion Factors The following Reference Conversion Table provides conversion factors for converting between units. To use this table: Move across the table to find the starting unit of measure. Move down the table to reach the ending unit of measure.
  • Page 329: Extended Features Map

    Extended Features Map ® Page E-2 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 330 Extended Features Map (continued) ® p/n 18603-001, Rev. J 1200, 1150, and 1100 Ventilator Service Manual Page E-3...
  • Page 331 Extended Features Map (continued) ® Page E-4 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 332: External Accessories Screw Location, Type And Length

    External Accessories Screw Location, Type and Length Damage to internal components of the ventilator can result if the wrong length mounting screws are used when permanently (or temporarily) removing or exchanging external accessories. Refer to the information on the following page to determine the appropriate external accessories mounting screws or accessories replacement mounting screws location, type ®...
  • Page 333 ® Page E-6 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 334: Parts, Replacement

    Parts, Replacement The following Replacement Parts table provides the name and part number of all parts ® potentially used in the servicing of the LTV 1200, 1150, and 1100 ventilator. Replacement Part Name Part Number Accumulator 18144-001 Adhesive, Loctite 4541 10773 Adhesive, RTV Silicon 10122...
  • Page 335 Replacement Part Name Part Number Label, Battery Replacement 18889-001 Label, Power 17420-001 Label, Transducer 10115 Label, Stepping Motor Connector 11322 Lubricant, Silicone Gel 10123 Mounting Block, earlier version LTM™ 11146 Mounting Bracket, earlier version LTM™ 11125 Mounting Block, current version LTM™ 17917-001 Mounting Bracket, current version LTM™...
  • Page 336 Replacement Part Name Part Number Screw, 1 7/8” Black colored Pan-head 10918B Screw, 1/4” Flat-head 10430 Screw, 1/4” Pan-head 10435 Screw, 3/16” Pan-head 14372 Screw, 3/8” Pan-head 18296-001 Screw, 3/4” Pan-head 10500 Screw, ½” Flat-head 10338 Screw, 1/4” Green colored Pan-head 10435G Screw, 5/8”...
  • Page 337: Settings, Dip Switch

    Settings, Dip Switch The following Dip Switch Settings table provides DIP Switch factory set default information. Dip Switch Default Function Setting Force all LEDs to turn on. Use in conjunction with dip switch #5. Intended for factory use only Force O pressure to 50 psi regardless of inlet pressure.
  • Page 338: Settings, Front Panel Controls And Extended Features

    Settings, Front Panel Controls and Extended Features The following tables provide factory default information. Control - Default Control - Default Breath Rate - 12 bpm % (O Flush) - 21% Control Lock - On PEEP - 0 cmH Data Display Scrolling - Auto-On Pressure Control - 1 cmH High Pressure Limit - 20 cmH Pressure Support - 1 cmH...
  • Page 339: Tools, Required

    Tools, Required ® Service, Calibration and Testing require instrumentation tools or test equipment with documented calibrated traceability to NIST standards. The instrumentation and test equipment must also have sufficient accuracies to attain industry standard performance levels. Instrumentation accuracy is typically specified “percent of full scale” or “percent of reading”. The required minimum accuracy levels are typically the specification tolerance divided by four (4).
  • Page 340 Flow rate measurements 10 Lpm When measuring a flow rate of 10 Lpm with a tolerance of ± 1.5 Lpm, the minimum accuracy level for instrumentation would be 1.5 Lpm / 4 = 0.375 Lpm. If the instrument accuracy is 2% of reading, at 10 Lpm the resulting accuracy is 10 Lpm x 2% = 0.2 Lpm.
  • Page 341 When measuring oxygen blending of 60% with a tolerance of ± 5%, the resulting minimum accuracy level for instrumentation would be 5% / 4 = 1.25%. If the instrument accuracy is 2% of reading, at 60% the resulting accuracy is 60% x 2% = 1.2%.
  • Page 342 ® The following unique tools are required to perform various procedures on the LTV 1200, 1150, and the 1100 ventilator and are available from CareFusion separately, or as part of the Maintenance and Calibration Kit, P/N 11566;  Calibration Syringe Assembly P/N 11471 ...
  • Page 343: Transducer Calibration, Acceptable A/D Counts

    128-3968 Flow Valve Differential (FVd) @ 15 cmH 1915 – 2340 (For Software Version < 05.04) 1915 – 2521 (For Software Version ≥ 05.04) Ambient (LTV 1200/1150) 40 – 328 Ambient (LTV 1100) 61 – 354 PEEP Pilot Pressure (PPP)
  • Page 344 Torque Values ® The following Torque Values Table provides torque values used in the assembly of the LTV 1200, 1150, and 1100 ventilator. In-oz Assy./Part to be Torqued Torque Value Accumulator to Turbine Manifold Spacer 60 in-oz (0.42 Nm) Analog Board mounting screws (2) 60 in-oz (0.42 Nm) Back Panel side mounting screws (6) 20 in-oz (0.14 Nm)
  • Page 345 ® Page E-18 1200, 1150, and 1100 Ventilator Service Manual p/n 18603-001, Rev. J...
  • Page 346 Appendix F - I NDEX RESET · 6-31, 7-36 SBT · 7-38 1  SBT < f · 7-38 SBT < f/Vt · 7-39 SBT > f · 7-38 12 hour burn-in SBT > f/Vt · 7-39 test · 9-32 Vent Inop · 2-10, 9-2 , 9-5, 9-13, 9-14, A-10 worksheet ·...
  • Page 347 calibration · 6-6 control settings · 6-36 acceptable A/D counts · E-16 EEPROM · 6-36 airway pressure · 6-8 events · 6-36 airway pressure transducers · 6-10 menu · 6-36 CAL FAIL · 6-10, 6-13, 6-16, 6-19, 6-21 component clearing · see clear, calibration removal ·...
  • Page 348 FLOW DIFF · 6-10, 6-11 inlet · 4-3 LEAK · 2-8 internal inlet · 8-85 LOCKED · 7-4 inlet · 4-4, 8-104 MOTOR DRIVE · 6-19 oxygen blender · 8-104 O2 INLET · 6-18 wet or damp · 4-5, 8-65 PPP ·...
  • Page 349 HW FAULT · 4-2, 7-22, 7-34, 8-65 map, extended features · E-2 measured T/S value · See turbine speed turbine speed · 6-22 I  mechanical shock · 6-26 membrane switch panel · 8-77 align · 8-78 inlet filter · 4-3 INSP HOLD ·...
  • Page 350 worksheet · 9-27 connector · 8-75, 8-78 pigtail cable assembly · 8-111 specifications · A-1 pisco connector · 8-59, 8-60, 8-66, 8-132 step test · 6-30 PM · see preventative maintenance Stepper Motor Calibration · 6-20 PNT ASSIST · 7-37, 7-38 sterilizing ·...
  • Page 351 instrumentation and test equipment solenoid mount and subassemblies requirements · E-12 installation · 8-45 PCB separator · 8-107 solenoid mount tube routing table · 8-36 required · E-15 tee & elbow subassembly · 8-39 required minimum test equipment accuracy tee subassembly · 8-41 levels ·...
  • Page 352 warranty · ii X  watchdog test · 6-31 XDCR FAULT · 7-7, 7-12, 7-33, 7-41 watchdog test · 9-24 XDCR ZERO · 7-8, 7-10, 7-13, 7-15, 7-19, 7- worksheets 12 hr burn-in · 9-34 battery checkout · see power checkout worksheet, see power checkout Z ...

This manual is also suitable for:

Ltv 1150Ltv 1100

Table of Contents