Boots, Temporary Removal

Parts required for replacement:
None
Replace if damaged:
Finish Washers (6) P/N 10191, or
19119-002
To temporarily remove the upper protective boot
1)
Carefully support the disconnected ventilator in an upright position on a clean, dry
surface.
2)
Using a Philips screwdriver, remove the two flat-head mounting screws and finish
washers in the legs of the upper boot and the two flat-head mounting screws and finish
washers in the sides of the upper boot, as indicated in the illustration.
3)
Retain the screws and finish washers for reuse when the upper boot is reinstalled.
4)
Remove the upper boot.
28
Contained in CareFusion Replacement Screws kit, P/N 11149.
29
See Appendix E - Reference Information, External Accessories Screw Location, Type and Length for
additional information.
p/n 18603-001, Rev. J
28
®
LTV
1200, 1150, and 1100 Ventilator Service Manual
Tools required:
Phillips screwdriver with
torque meter
29
:
Page 8-5